then pull apart the connectors. 

IV. 

pH sensor calibration 

 

IMPORTANT: Exposure of an ammonium sensor to the pH buffers will cause extremely long 

ammonium calibration times.  If an ammonium sensor is installed, cover it with a tight rubber seal 

or remove it and insert a port plug before calibrating the pH sensor. 

A. 

Press 

Cal

. Select 

None

. Highlight 

ISE (pH)

 and press 

enter

. The message line will show 

the instrument is “Ready for point 1”. The pH calibration allows up to six calibration 

points. 

 

 

 

B. 

Place the sensor in a pH 7 buffer solution. The instrument should automatically recognize 

the buffer value and display it at the top of the calibration screen. If the calibration value 

is incorrect, the auto buffer recognition setting in the Sensor Setup menu may be 

incorrect. If necessary, highlight the Calibration Value and press enter to input the correct 

buffer value. After 5 minutes and once the pH and temperature readings stabilize, 

highlight 

Accept

 

Calibration

 and press 

enter

 to accept the first calibration point. The 

message line will then display “Ready for point 2”. 

C. 

Rinse and place the sensor in the second buffer solution. The instrument should 

automatically recognize the second buffer value and display it at the top of the screen. If 

necessary, highlight the Calibration Value and press enter to input the correct buffer 

value. After 5 minutes and once the pH and temperature readings stabilize, highlight 

Accept Calibration

 and press 

enter

 to confirm the second calibration point. The 

message line will then display ‘Ready for point 3” and you can continue with the 3

rd

 

calibration point if desired. If you do not wish to perform a 3rd calibration point, press 

Cal

 to complete the calibration. 

V. 

Ammonium sensor calibration 

 

The calibration procedures for ammonium, nitrate, or chloride are similar to pH. The only 

differences are the calibration solutions.  If the ammonium sensor was capped in order to calibrate 

the pH sensor, remove the cap before beginning the ammonium sensor calibration. 

A. 

Place the proper amount of 1 mg/L Ammonium standard into the pre-rinsed calibration 

cup. Carefully immerse the sensor into the solution. Allow at least 1 minute for 

temperature equilibration before proceeding. 

B. 

Press 

Cal

. Select 

None

. Highlight 

ISE (Ammonium)

 and press 

enter

. The message line 

will show the instrument is ready for the 1st calibration point.  The instrument will 

display the calibration value at the top of the screen. If necessary, highlight the 

Calibration value and press enter to input the correct value. After 5 minutes and once the 

ammonium and temperature readings stabilize, highlight 

Accept

 

Calibration

 and press 

enter

 to accept the first calibration point. The message line will then display “Ready for 

point 2”. 

C. 

Rinse the sensor with clean water, then dry it before placing it in the 100 mg/L 

calibration standard. Allow at least 5 minutes for temperature equilibration before 

Summary of Contents for 1001

Page 1: ...ntrol Centre 1219 Queen Street East Sault Ste Marie Ontario P6A 2E5 Canada INSTRUMENT OPERATING PROCEDURE INSTRUMENT Multiparameter Water Quality Instrument MODEL Pro Plus Meter 6050000 Quatro Cable 605790 Probes temperature conductivity 5560 pH 1001 or 1001A DO 2003 ammonium 1004 MANUFACTURER Yellow Springs Instrument Co Inc YSI PRECAUTIONS POTENTIAL INTERFERENCES General Greatest accuracy is ach...

Page 2: ...ulated from ammonium sensor measurements Exposure to the high ionic content of pH buffers can cause a significant but temporary drift in the ammonium sensors Therefore when calibrating the pH sensor remove ammonium sensor from the cable bulkhead and plug the port After pH calibration is complete replace the ISE sensors and proceed with ammonium calibration Alternatively the ammonium sensor may be ...

Page 3: ... screen Use to calibrate all parameters except temperature File Opens File Menu from any screen Use to view data and GLP files set up site and folder lists and delete data Backlight Press to turn the instrument backlight on and off and to adjust the display contrast when pressed with the left or right arrow key Right Arrow Use to navigate right in alpha numeric entry screens Can be pressed simulta...

Page 4: ... B Instrument software setup 1 Press the Power key to turn the instrument on The instrument will briefly display the splash screen with the YSI logo then go directly to the main run screen The first time the instrument is powered up or if the instrument has had a battery change with batteries removed for more than 2 minutes you will need to set the date and time a Highlight Date Time from the Syst...

Page 5: ...ress enter 2 Highlight pH and press enter to enable it 3 Highlight pH mV and press enter to enable it 4 Press to return to the previous menu c Ammonium sensor 1 Highlight ISE2 NH4 Press enter 2 Highlight NH4 N mg l and press enter to enable it 3 Highlight NH4 mV and press enter to disable it 4 Highlight NH3 N mg l and press enter to enable it 5 Press to return to the previous menu d Dissolved oxyg...

Page 6: ...etely without feeling any resistance Damage to your cable sensor may occur if you force the parts together e Once completely installed the nut will seat flat against the bulkhead At this point use the included tool to turn the nut an additional to turn DO NOT overtighten Refer to the sensor installation sheet that is included with the conductivity temperature sensor for detailed instructions 4 Ins...

Page 7: ... clockwise to engage threads and finger tighten Do not use a tool This connection is waterproof Refer to the sensor installation sheet included with each sensor for detailed instructions 7 Install the 2003 DO sensor membrane as follows a Change membrane and oxygen probe electrolyte solution every 30 days b Remove the metal sensor guard 3 If installing a membrane for the first time remove the prote...

Page 8: ... buffer value After 5 minutes and once the pH and temperature readings stabilize highlight Accept Calibration and press enter to confirm the second calibration point The message line will then display Ready for point 3 and you can continue with the 3rd calibration point if desired If you do not wish to perform a 3rd calibration point press Cal to complete the calibration V Ammonium sensor calibrat...

Page 9: ...he instrument will use the internal barometer during calibration and will display this value in brackets at the top of the display Highlight Barometer and press enter to adjust it if needed If the barometer reading is incorrect it is recommended that you calibrate the barometer Note the barometer should be reading true barometric pressure see Barometer section of Instrument Manual for more informa...

Page 10: ...low the sensor sufficient time to stabilize 4 pH sensor The key to pH sensor storage short or long term is to make certain that the sensor does not dry out Sensors which have been allowed to dry out due to improper storage procedures may be irreparably damaged by the dehydration and will require replacement You can try to rehydrate the sensor by soaking it preferably overnight in a potassium chlor...

Page 11: ...Jessica M Barber Digitally signed by JESSICA BARBER Date 2020 03 10 11 22 25 04 00 ...

Reviews: