WILDLIFE 1001 Quick Start Manual Download Page 2

IOP 019A.1 

Effective Date: 1/11/2017 

Page 2 of 11 

 

 

 

Dissolved oxygen

 – 

-

 

Chlorine, sulfur dioxide, nitric oxide, and nitrous oxide can affect readings by behaving like 

oxygen at the membrane. 

-

 

Static electricity may affect readings. 

-

 

Gold Cathode - The gold cathode on the sensor must be textured properly.  It may become 

tarnished or plated with silver after extended periods of use. 

-

 

Bubbles under the membrane cap will affect readings. 

-

 

Excessive water in calibration/storage sleeve will affect calibration. 

 

pH

 – 

-

 

Do not measure pH in metal containers.

 

-

 

Water currents in streams can cause unstable readings. 

-

 

pH measurements will drift if the sample is not stirred. 

-

 

Do not expose the meter or probe to freezing temperatures. 

-

 

High humidity and cold temperatures may affect meter operation. 

-

 

Be sure that the temperature sensor is immersed in the sample. 

-

 

Buffers degrade from exposure to light - change daily 

 

Ammonium

 – 

-

 

Ammonium sensors should only be used at DEPTHS OF LESS THAN 55 FEET (17 

METERS). Use of the sensors at greater depths is likely to permanently damage the sensor 

membrane. 

-

 

Ammonium sensors should only be used in FRESHWATER. 

-

 

Temperature and pH sensors must be installed, calibrated and functioning correctly in order 

for accurate ammonia values to be calculated from ammonium sensor measurements. 

-

 

Exposure to the high ionic content of pH buffers can cause a significant, but temporary, drift 

in the ammonium sensors. Therefore, when calibrating the pH sensor, remove ammonium 

sensor from the cable bulkhead and plug the port. After pH calibration is complete, replace 

the ISE sensors and proceed with ammonium calibration.  Alternatively, the ammonium 

sensor may be covered with a tightly fitting rubber cap during pH calibration.  Failure to 

follow this protocol will result in ammonium calibration times of SEVERAL HOURS. 

 

SAFETY

 

 

No special precautions 

 

PROCEDURES: 

 

I. 

Configuration – The YSI Pro Plus instrument is capable of simultaneous measurement of multiple 

water chemistry parameters including temperature, conductivity, pH, dissolved oxygen, 

ammonium, chloride, and nitrate.    This IOP is limited to the parameters typically measured by 

the SLCP (temperature, pH, dissolved oxygen, and ammonia).  Information on additional 

capabilities of the Pro Plus instrument are available in the User Manual and Calibration Tips 

documents available from YSI. 

A. 

Instrument overview 

1. 

Meter – The YSI Pro Plus handheld meter includes internal hardware and 

software for calculation of parameters, a display screen, and a keypad for data 

entry and screen navigation.  The meter includes an internal barometer. 

2. 

Cable – The instrument cable connects the meter to the sensors.  The SLCP uses 

the quatro cable which has a four sensor capacity. 

Summary of Contents for 1001

Page 1: ...ntrol Centre 1219 Queen Street East Sault Ste Marie Ontario P6A 2E5 Canada INSTRUMENT OPERATING PROCEDURE INSTRUMENT Multiparameter Water Quality Instrument MODEL Pro Plus Meter 6050000 Quatro Cable 605790 Probes temperature conductivity 5560 pH 1001 or 1001A DO 2003 ammonium 1004 MANUFACTURER Yellow Springs Instrument Co Inc YSI PRECAUTIONS POTENTIAL INTERFERENCES General Greatest accuracy is ach...

Page 2: ...ulated from ammonium sensor measurements Exposure to the high ionic content of pH buffers can cause a significant but temporary drift in the ammonium sensors Therefore when calibrating the pH sensor remove ammonium sensor from the cable bulkhead and plug the port After pH calibration is complete replace the ISE sensors and proceed with ammonium calibration Alternatively the ammonium sensor may be ...

Page 3: ... screen Use to calibrate all parameters except temperature File Opens File Menu from any screen Use to view data and GLP files set up site and folder lists and delete data Backlight Press to turn the instrument backlight on and off and to adjust the display contrast when pressed with the left or right arrow key Right Arrow Use to navigate right in alpha numeric entry screens Can be pressed simulta...

Page 4: ... B Instrument software setup 1 Press the Power key to turn the instrument on The instrument will briefly display the splash screen with the YSI logo then go directly to the main run screen The first time the instrument is powered up or if the instrument has had a battery change with batteries removed for more than 2 minutes you will need to set the date and time a Highlight Date Time from the Syst...

Page 5: ...ress enter 2 Highlight pH and press enter to enable it 3 Highlight pH mV and press enter to enable it 4 Press to return to the previous menu c Ammonium sensor 1 Highlight ISE2 NH4 Press enter 2 Highlight NH4 N mg l and press enter to enable it 3 Highlight NH4 mV and press enter to disable it 4 Highlight NH3 N mg l and press enter to enable it 5 Press to return to the previous menu d Dissolved oxyg...

Page 6: ...etely without feeling any resistance Damage to your cable sensor may occur if you force the parts together e Once completely installed the nut will seat flat against the bulkhead At this point use the included tool to turn the nut an additional to turn DO NOT overtighten Refer to the sensor installation sheet that is included with the conductivity temperature sensor for detailed instructions 4 Ins...

Page 7: ... clockwise to engage threads and finger tighten Do not use a tool This connection is waterproof Refer to the sensor installation sheet included with each sensor for detailed instructions 7 Install the 2003 DO sensor membrane as follows a Change membrane and oxygen probe electrolyte solution every 30 days b Remove the metal sensor guard 3 If installing a membrane for the first time remove the prote...

Page 8: ... buffer value After 5 minutes and once the pH and temperature readings stabilize highlight Accept Calibration and press enter to confirm the second calibration point The message line will then display Ready for point 3 and you can continue with the 3rd calibration point if desired If you do not wish to perform a 3rd calibration point press Cal to complete the calibration V Ammonium sensor calibrat...

Page 9: ...he instrument will use the internal barometer during calibration and will display this value in brackets at the top of the display Highlight Barometer and press enter to adjust it if needed If the barometer reading is incorrect it is recommended that you calibrate the barometer Note the barometer should be reading true barometric pressure see Barometer section of Instrument Manual for more informa...

Page 10: ...low the sensor sufficient time to stabilize 4 pH sensor The key to pH sensor storage short or long term is to make certain that the sensor does not dry out Sensors which have been allowed to dry out due to improper storage procedures may be irreparably damaged by the dehydration and will require replacement You can try to rehydrate the sensor by soaking it preferably overnight in a potassium chlor...

Page 11: ...Jessica M Barber Digitally signed by JESSICA BARBER Date 2020 03 10 11 22 25 04 00 ...

Reviews: