WILDLIFE 1001 Quick Start Manual Download Page 10

IOP 019A.1 

Effective Date: 1/11/2017 

Page 10 of 11 

 

 

 A. 

Short term storage 

The cable assembly is supplied with a sensor storage container, or sleeve, that attaches to 

the cable. The container is used for short-term storage (less than 7 days). Be sure to keep 

a small amount of moisture (tap water) in the container during storage. This is done to 

maintain a 100% saturated air environment which is ideal for short-term sensor storage. 

The sensors should not be submersed in water. The intent is to create a humid air storage 

environment. 

B. 

Long term storage 

1. 

Temperature sensor - No special storage is required. The temperature sensor can 

be stored dry or wet as long as solutions in contact with the thermistor are not 

corrosive (for example, chlorine bleach). 

2. 

Conductivity sensor - No special storage is required. Sensors can be stored dry 

or wet as long as solutions in contact with conductivity electrodes are not 

corrosive (for example, chlorine bleach). However, it is recommended that the 

sensor be cleaned with the provided brush prior to and after long term storage. 

3. 

Dissolved oxygen sensor - Dissolved oxygen sensors should be stored in a dry 

state for long term storage. First, remove the membrane cap and thoroughly 

rinse the sensor with clean water. Next, either blow it dry with compressed air or 

allow to air dry completely. Install a clean, dry new membrane cap over the 

sensor to keep it dry and to protect the electrodes. After storing the sensor for a 

long period of time, it is necessary to “condition” the sensor by putting a new 

membrane with electrolyte solution on the sensor and then turning the 

instrument on to allow the sensor sufficient time to stabilize. 

4. 

pH sensor - The key to pH sensor storage, short or long-term, is to make certain 

that the sensor does not dry out. Sensors which have been allowed to dry out due 

to improper storage procedures may be irreparably damaged by the dehydration 

and will require replacement. You can try to rehydrate the sensor by soaking it 

(preferably overnight) in a potassium chloride solution or a pH 4 buffer before 

attempting to calibrate. To store the sensor, remove it from the cable and seal the 

vacant port with a port plug. Fill the original shipping/storage vessel (plastic 

boot or bottle) with buffer 4 solution and then submerge the sensor into the 

solution. The sensor should remain submerged in the solution during the storage 

period; therefore, make certain that the vessel is sealed to prevent evaporation 

and periodically check the vessel to ensure the sensor does not dry out. 

5. 

Ammonium sensor - The key to ISE sensor storage, short or long-term, is to 

make certain that the sensor does not dry out. Sensor junctions that have been 

allowed to dry out due to improper storage procedures may be irreparably 

damaged by the dehydration and will require replacement. You can attempt to 

rehydrate the sensor by soaking it (preferably overnight) in the sensor’s high 

calibration solution before attempting to calibrate. The recommended storage of 

these sensors is in moist air. Remove the sensor from the cable and seal the 

vacant port with the provided plug. Place the sensor in its original shipping 

storage vessel (plastic boot or bottle) with a small amount of tap water or its 

high calibration standard. The vessel should remain a saturated air environment. 

The sensor only needs to be kept in moist air, not submerged. Make certain that 

the vessel is sealed to prevent evaporation. 

MAINTENANCE: 

Summary of Contents for 1001

Page 1: ...ntrol Centre 1219 Queen Street East Sault Ste Marie Ontario P6A 2E5 Canada INSTRUMENT OPERATING PROCEDURE INSTRUMENT Multiparameter Water Quality Instrument MODEL Pro Plus Meter 6050000 Quatro Cable 605790 Probes temperature conductivity 5560 pH 1001 or 1001A DO 2003 ammonium 1004 MANUFACTURER Yellow Springs Instrument Co Inc YSI PRECAUTIONS POTENTIAL INTERFERENCES General Greatest accuracy is ach...

Page 2: ...ulated from ammonium sensor measurements Exposure to the high ionic content of pH buffers can cause a significant but temporary drift in the ammonium sensors Therefore when calibrating the pH sensor remove ammonium sensor from the cable bulkhead and plug the port After pH calibration is complete replace the ISE sensors and proceed with ammonium calibration Alternatively the ammonium sensor may be ...

Page 3: ... screen Use to calibrate all parameters except temperature File Opens File Menu from any screen Use to view data and GLP files set up site and folder lists and delete data Backlight Press to turn the instrument backlight on and off and to adjust the display contrast when pressed with the left or right arrow key Right Arrow Use to navigate right in alpha numeric entry screens Can be pressed simulta...

Page 4: ... B Instrument software setup 1 Press the Power key to turn the instrument on The instrument will briefly display the splash screen with the YSI logo then go directly to the main run screen The first time the instrument is powered up or if the instrument has had a battery change with batteries removed for more than 2 minutes you will need to set the date and time a Highlight Date Time from the Syst...

Page 5: ...ress enter 2 Highlight pH and press enter to enable it 3 Highlight pH mV and press enter to enable it 4 Press to return to the previous menu c Ammonium sensor 1 Highlight ISE2 NH4 Press enter 2 Highlight NH4 N mg l and press enter to enable it 3 Highlight NH4 mV and press enter to disable it 4 Highlight NH3 N mg l and press enter to enable it 5 Press to return to the previous menu d Dissolved oxyg...

Page 6: ...etely without feeling any resistance Damage to your cable sensor may occur if you force the parts together e Once completely installed the nut will seat flat against the bulkhead At this point use the included tool to turn the nut an additional to turn DO NOT overtighten Refer to the sensor installation sheet that is included with the conductivity temperature sensor for detailed instructions 4 Ins...

Page 7: ... clockwise to engage threads and finger tighten Do not use a tool This connection is waterproof Refer to the sensor installation sheet included with each sensor for detailed instructions 7 Install the 2003 DO sensor membrane as follows a Change membrane and oxygen probe electrolyte solution every 30 days b Remove the metal sensor guard 3 If installing a membrane for the first time remove the prote...

Page 8: ... buffer value After 5 minutes and once the pH and temperature readings stabilize highlight Accept Calibration and press enter to confirm the second calibration point The message line will then display Ready for point 3 and you can continue with the 3rd calibration point if desired If you do not wish to perform a 3rd calibration point press Cal to complete the calibration V Ammonium sensor calibrat...

Page 9: ...he instrument will use the internal barometer during calibration and will display this value in brackets at the top of the display Highlight Barometer and press enter to adjust it if needed If the barometer reading is incorrect it is recommended that you calibrate the barometer Note the barometer should be reading true barometric pressure see Barometer section of Instrument Manual for more informa...

Page 10: ...low the sensor sufficient time to stabilize 4 pH sensor The key to pH sensor storage short or long term is to make certain that the sensor does not dry out Sensors which have been allowed to dry out due to improper storage procedures may be irreparably damaged by the dehydration and will require replacement You can try to rehydrate the sensor by soaking it preferably overnight in a potassium chlor...

Page 11: ...Jessica M Barber Digitally signed by JESSICA BARBER Date 2020 03 10 11 22 25 04 00 ...

Reviews: