In the event of any faults, first check whether the flow switch is mounted correctly, mechanically and electrically.
If complaint is unjustified, the handling costs will be charged.
Error display
Via the instrument's display internal errors are shown.
The following table shows the error codes and their meaning.
Error code
Description
ATT1
On changing the switch point, the system automatically reduces the reset point.
ATT3
Password entered for menu access is incorrect
ATT4
TAG cannot be shown in the display (e.g. special characters)
ERR
Internal error
OL F
Measuring range flow exceeded ≥ 10 % (display flashing)
OL T
Measuring range temperature exceeded ≥ 5 °C (display flashing)
UL T
Below measuring range temperature ≤ 5 °C (display flashing)
Acknowledgement of an error display by pressing the “Enter” key.
Fault
Possible cause
Measure
No output signal
Cable break
Check the continuity
No output signal/cable break
Mechanical load too high
Replace probe with a suitable design
No output signal
No/wrong power supply
Rectify the power supply
No/wrong output signal
Wiring error
Observe the pin assignment
Wrong output signal
Process temperature out of measuring
range; sensor drift caused by overtemper
-
ature
Check temperature range
Probe break/short circuit
Send the instrument to the manufacturer
6. Faults
Summary of Contents for FSD-3
Page 38: ...38 WIKA operating instructions flow switch model FSD 3 EN 14153887 01 09 2016 EN DE FR...
Page 74: ...74 WIKA Betriebsanleitung Str mungsschalter Typ FSD 3 DE 14153887 01 09 2016 EN DE FR...
Page 110: ...110 WIKA mode d emploi pour le capteur de d bit type FSD 3 FR 14153887 01 09 2016 EN DE FR...
Page 111: ...111 WIKA mode d emploi pour le capteur de d bit type FSD 3 FR 14153887 01 09 2016 EN DE FR...