5.1 Mechanical mounting
■
The sealing faces at the instrument and at the
measuring location have to be undamaged and
clean.
■
When screwing in the instrument, the force
required to do this must not be applied through
the case, but only through the spanner flats
provided for this purpose and using a suitable
tool.
The correct torque depends on the dimensions of the process connection and the sealing used
(form/material).
Do not overwind the process connection, do not exceed the permissible torques.
■
When screwing in, do not cross the threads.
■
For information on tapped holes and welding sockets, see Technical information IN 00.14 at
www.wika.com.