20
WIKA operating instructions, model CTD9100-1100
14030725.02 08/2019 EN/DE
EN
12. Specifications
Model CTD9100-1100
Indication
Display
2-line LC display 20 ch x lines (3.2 x 5.5 in) with backlighting
Temperature range
200 … 1,100 °C [392 ... 2,012 °F]
Accuracy
1)
±3 K
Stability
2)
±0,4 K at 1,000 °C [2,012 °F]
Display resolution
0.01 °C up to 999 °C, then 0.1 °C [0.01 up to 1,830 °F, then 0.1]
Temperature distribution
Axial homogeneity
3)
dependent on temperature, temperature probes and their quantity
Radial homogeneity
4)
dependent on temperature, temperature probes and their quantity
Temperature control
Heating time
approx. 50 min (approx. 18 °C/min) from 20 °C to 900 °C [from 68 °F to 1,652 °F]
Cooling time
approx. 150 min from 1,100 °C to 200 °C [from 2,012 °F to 392 °F]
Stabilisation time
5)
dependent on temperature and temperature probe
Temperature block
Dimensions (D x H)
Ø 44 x 300 mm [Ø 1.73 x 11.81 in]
Block depth
175 mm [6.89 in]
Insert
Immersion depth
155 mm [6.10 in]
Insert dimensions
Ø 42.5 x 175 mm [Ø 1.67 x 6.89 in]
Insulator dimensions
Ø 42.5 x 65 mm [Ø 1.67 x 2.56 in]
Insert material
Ceramic
Voltage supply
Power supply
AC 230 V, 50/60 Hz or AC 110 V, 50/60 Hz
Power consumption
950 VA
Fuse
6.3 A slow blow fuse
Power cord
AC 230 V; for Europe
Communication
Interface
RS-232
Case
Dimensions (W x D x H)
170 x 390 x 330 mm [6.69 x 15.35 x 12.99 in]
Weight
12 kg [26.46 lbs]
1) Is defined as the measuring deviation between the measured value and the reference value.
2) Maximum temperature difference at a stable temperature over 30 minutes.
3) Maximum temperature difference at 40 mm [1.57 in] above the bottom.
4) Maximum temperature difference between the bores (all thermometers inserted to the same depth).
5) Time before reaching a stable value.
The measurement uncertainty is defined as the total measurement uncertainty (k = 2), which contains the following shares:
accuracy, measurement uncertainty of reference, stability and homogeneity.
Certificates
Certificate
Calibration
Standard: 3.1 calibration certificate per DIN EN 10204
Option: DKD/DAkkS calibration certificate
Recommended recalibration interval
1 year (dependent on conditions of use)
Approvals and certificates, see website
For further specifications see WIKA data sheet CT 41.29 and the order documentation.
12. Specifications