WIKA CTD9100-1100 Operating Instructions Manual Download Page 12

12

WIKA operating instructions, model CTD9100-1100

14030725.02 08/2019 EN/DE

EN

After approx. 5 secs, the initialisation will be complete and 

the 

calibration mode

 will automatically be displayed.

The built-in heating or cooling elements will temper the metal 

block automatically from room temperature to the controller’s 

set temperature.

6.3  Start-up procedure

If the calibrator is not used for a long time, because of the 

materials used (magnesium oxide), moisture can penetrate 

the heating elements.

Following transport or storage of the calibrator in moist 

environments, the heating elements must be taken to higher 

temperatures slowly.

During the drying process,it should be assumed that the 

calibrator has not yet reached the required insulation voltage 

for protection class I. The start-up set point is T

anf

 

= 400 °C 

[752 °F] with a holding time of T

n

 = 15 min.

6.4  Temperature switch test

It is possible to control the intervention temperature of a 

temperature switch by the 

“Switch-Test”

 function.

1.  Insert the temperature switch in the most suitable hole of 

the insert.

2.  Connect the temperature switch electrical output cables to 

the terminals.

3.  Turn the calibrator on.

4.  Set the intervention temperature and check the release by 

the lighting of the indication light.

5.  The temperature switch release values are recorded. In 

order to display the recorded value, refer to the procedure 

explained in chapter 7.2.2 “First level - General settings” 

till 

“SW ON - SW OFF”

.

 

Press the button 

[▲]

 and 

[▼]

 at the same time in order 

to reset the 

“SW ON - SW OFF”

 values.

6.  Refer to item 7.2 “Programming (Main menu)” or 

7.2.2 “First level - General settings” to set ramps in order 

to reduce the ascent rate and descent rate of temperature 

to test the temperature switch

6.5  Testing of temperature sensors

To test temperature sensors, connect a separate 

temperature-measuring instrument to the test item. By 

comparing the temperature displayed on the external 

measuring instrument with the reference temperature, 

there is evidence of the status of the test item. This ensured 

that the test item requires a short time until it reaches the 

temperature of the metal block.

6.6  Positioning of the probe

The sleeve is made in CERAMIC and it has different holes 

on the inside to make it possible to calibrate various types of 

probes. The function of this block is to make the temperature 

uniform. It is also possible to calibrate probes of different 

length by adapting the depth of the holes. A customized 

sleeve can be ordered at the manufacturer.

A customised sleeve can be ordered at the 

manufacturer.

After the furnace generally been installed, the insert and 

the ceramic tube insulation may be inserted. Carefully insert 

the sleeve and the upper insulation into the Ceramic tube. 

Prevent dirt or other foreign materials between the block and 

the ceramic tube.

The insert replacement tool consists of a stainless steel 

rod with a threaded end, which is screwed into the top of 

the insert. The sleeve has to be inserted in this way that 

the grooves are directly adapted to the control and cutout 

thermometer. Insert the ceramic tube insulation on the top of 

the sleeve. Centre the holes of the upper insulation with the 

holes of the insert.

Whenever the calibrator has to be moved, 

remove the sleeve and the ceramic tube in order 

to prevent breakage.

To obtain the best results for the calibration, following advises 

are important:

 

The diameter of the hole in the calibration block is at least 

1 mm bigger than the diameter of the probe (see Fig.1).

 

The sensitive element of the thermometer is in the best 

position, if the probe is inserted down to the bottom (see 

Fig. 2).

When the probes length is shorter than the length of the hole 

in the sleeve, the reference sensor should be on the same 

height as the test item. Moreover the thermometers have to 

be as close together as possible (see Fig. 3).

 

The temperature difference is proportional to 

the difference between the diameter of the 

probe and the diameter of the hole

 

The test item has to be inserted before the 

set temperature is reached, otherwise this 

can cause instabilities or a breakage of the 

sensitive element.

6. Commissioning, operation

Summary of Contents for CTD9100-1100

Page 1: ...Operating instructions Betriebsanleitung EN DE Temperature dry well calibrator model CTD9100 1100 Temperatur Blockkalibrator Typ CTD9100 1100 Temperature dry well calibrator model CTD9100 1100...

Page 2: ...nguages can be found at www wika com 04 2011 WIKA Alexander Wiegand SE Co KG All rights reserved Alle Rechte vorbehalten WIKA is a registered trademark in various countries WIKA ist eine gesch tzte Ma...

Page 3: ...oning operation 11 6 1 Operating position 11 6 2 Switching on the calibrator 11 6 3 Start up procedure 12 6 4 Temperature switch test 12 6 5 Testing of temperature sensors 12 6 6 Positioning of the pr...

Page 4: ...icinity of the instrument and readily accessible to skilled personnel at any time Pass the operating instructions onto the next operator or owner of the instrument Skilled personnel must have carefull...

Page 5: ...r 12 Specifications The appropriate instrument should be selected depending on the application it should be connected correctly tests carried out and all components serviced This instrument is not per...

Page 6: ...instrument If the calibrator is used for purposes not expressly specified in these operating instructions additional safety measures have to be taken The Processor regulator has been configured in fac...

Page 7: ...ific regulations current standards and directives are capable of carrying out the work described and independently recognising potential hazards Special operating conditions require further appropriat...

Page 8: ...art contains the complete electronic unit for controlling the reference temperature Solid state relays SSR are used to control the heating and cooling elements Constant use at extreme temperatures red...

Page 9: ...creasing the value shown on display 2 Holding down the button will decrease the speed 7 Button Increasing the value shown on display 2 Holding down the button will increase the speed 8 Function menu b...

Page 10: ...cable are the scope of delivery The cable with the CEE polarized plug is designed to reduce the risk of electric shock polarizing the phase of the power If possible always use these cable Rx Tx 13 14...

Page 11: ...ases Potentially explosive environments flammable atmospheres Store the temperature dry well calibrator in its original packaging in a location that fulfils the conditions listed above 6 Commissioning...

Page 12: ...there is evidence of the status of the test item This ensured that the test item requires a short time until it reaches the temperature of the metal block 6 6 Positioning of the probe The sleeve is ma...

Page 13: ...button M This opens the main menu 2 With the button M to select the desired entry in the main menu see overview 3 With the button E confirm the selected menu item The calibrator has three function lev...

Page 14: ...14 WIKA operating instructions model CTD9100 1100 14030725 02 08 2019 EN DE EN 7 2 1 Menu structure parameter levels Please note Button F button M Menu structure 7 Operating of the calibrator...

Page 15: ...rs and on the lower the set point is displayed Set point 2 is reached acoustic signal and the ramp is paused A cooling ramp is set gradient must be negative and the set point 2 must be lower than the...

Page 16: ...then the buttons M and have to be pressed simultaneously Back over simultaneous pressing of the buttons M and or waiting 20 s Function Meaning Access key Access code Default 2 Baud rate Velocity of d...

Page 17: ...vironment caused by high temperatures Should a failure with high temperature may be present at the instrument Wear the requisite protective equipment CAUTION Physical injuries and damage to property a...

Page 18: ...al media in the dismounted instrument can result in a risk to personnel the environment and equipment Use the requisite protective equipment Carry out the cleaning process as described below 1 Cool do...

Page 19: ...medium Clean the instrument in order to protect persons and the environment from exposure to residual media 11 1 Dismounting WARNING Risk of burns During dismounting there is a risk of high temperatu...

Page 20: ...m 1 67 x 2 56 in Insert material Ceramic Voltage supply Power supply AC 230 V 50 60 Hz or AC 110 V 50 60 Hz Power consumption 950 VA Fuse 6 3 A slow blow fuse Power cord AC 230 V for Europe Communicat...

Page 21: ...21 WIKA operating instructions model CTD9100 1100 14030725 02 08 2019 EN DE EN 170 6 69 448 17 64 330 12 99 390 15 35 12 Specifications Top view Side view left Rear view Front view Dimensions in mm in...

Page 22: ...13 Accessories Accessories Order code Description CTX A K9 Transport case TB Power cord for EU EU for Switzerland CH for UK UK for USA Canada US Insert replacement tool RT Ordering information for you...

Page 23: ...Lagerung 31 6 Inbetriebnahme Betrieb 31 6 1 Betriebslage 31 6 2 Einschalten des Kalibrators 31 6 3 Anfahrprozedur 32 6 4 Temperaturschaltertest 32 6 5 Pr fen von Temperaturf hlern 32 6 6 Positionierun...

Page 24: ...he des Ger tes f r das Fachpersonal jederzeit zug nglich aufbewahrt werden Betriebsanlei tung an nachfolgende Benutzer oder Besitzer des Ger tes weitergeben 1 Allgemeines Das Fachpersonal muss die Be...

Page 25: ...rfen keinesfalls berschritten werden siehe Kapitel 12 Technische Daten Es ist je nach Anwendungsfall ein entsprechendes Instru ment auszuw hlen dieses korrekt anzuschlie en Tests durchzuf hren sowie...

Page 26: ...egelger t konzipiert Bei einem Gebrauch des Kalibrators f r nicht ausdr cklich in dieser Betriebsanleitung vorgesehene Anwendungen m ssen zus tzliche Schutzma nahmen getroffen werden Der elektronische...

Page 27: ...sowie Kenntnis der landesspezifischen Vorschriften geltenden Normen und Richtlinien in der Lage die beschriebenen Arbeiten auszuf hren und m gliche Gefahren selbstst ndig zu erkennen Spezielle Einsatz...

Page 28: ...lette Elektronik einheit zur Regelung der Referenztemperatur Zur Ansteuerung der Heizelemente werden Halbleiter Relais SSR verwendet Konstante Verwendung des Kalibrators an extre men Temperaturen verr...

Page 29: ...Wert welcher auf Anzeige 2 dargestellt wird Bei Halten der Taste wird die Geschwindigkeit verringert 7 Taste Erh hen des Wertes welcher auf Anzeige 2 darge stellt wird Bei Halten der Taste wird die G...

Page 30: ...in Euro Steckerkabel beinhaltet Das CEE Kabel wird mitgeliefert um einen elektrischen Schlag aufgrund Verpolung zu verhindern Wenn m glich immer diese Kabel benutzen Rx Tx 13 14 15 16 17 Bei Benutzung...

Page 31: ...ellen Ru Dampf Staub und korrosive Gase Explosionsgef hrdete Umgebung entz ndliche Atmosph ren Temperatur Blockkalibrator in der Originalverpackung an einem Ort lagern der die oben gelisteten Bedingun...

Page 32: ...nn eine Aussage ber den Zustand des Pr flings gemacht werden Hierbei darauf achten dass der Pr fling kurze Zeit ben tigt bis er die Temperatur des Metallblockes angenommen hat 6 6 Positionierung der T...

Page 33: ...7 Bedienung des Kalibrators 7 1 Einstellen einer tempor ren Soll Temperatur Sollwertmodus Einstellung der Soll Temperatur Dr cken der Taste erh ht den Sollwert Dr cken der Taste erniedrigt den Sollwe...

Page 34: ...34 WIKA Betriebsanleitung Typ CTD9100 1100 14030725 02 08 2019 EN DE DE 7 Bedienung des Kalibrators 7 2 1 Men struktur Parameterebenen Bitte beachten Taste F Taste M Men struktur...

Page 35: ...ert auf der zweiten Zeile des Displays angezeigt Ist der Sollwert 2 erreicht ert nt ein Signal und die Rampe wird automatisch pausiert Wenn eine runter laufende Rampe eingestellt wurde muss der Gradie...

Page 36: ...Code ber eingestellt wurde und danach gleichzeitig die Tasten M und gedr ckt werden Zur ck kommt man durch gleichzeitiges Dr cken der Tasten M und oder durch 20 s langes Warten Funktion Bedeutung Acce...

Page 37: ...hutzhandschuhe WARNUNG K rperverletzungen Sach und Umweltsch den durch hohe Temperaturen Am Ger t k nnen im Fehlerfall extreme Tempe raturen anliegen Notwendige Schutzausr stung tragen VORSICHT K rper...

Page 38: ...SICHT Besch digung des Ger tes Eine unsachgem e Reinigung f hrt zur Besch digung des Ger tes Keine aggressiven Reinigungsmittel verwen den Keine harten und spitzen Gegenst nde zur Reinigung verwenden...

Page 39: ...Um Sch den zu vermeiden 1 Das Ger t mit dem D mmmaterial in der Verpackung platzieren Zu allen Seiten der Transportverpackung gleichm ig d mmen 2 Wenn m glich einen Beutel mit Trocknungsmittel der Ver...

Page 40: ...7 x 2 56 in H lsenmaterial Keramik Spannungsversorgung Hilfsenergie AC 230 V 50 60 Hz oder AC 110 V 50 60 Hz Leistungsaufnahme 950 VA Sicherung 6 3 A tr ge Netzkabel AC 230 V f r Europa Kommunikation...

Page 41: ...anleitung Typ CTD9100 1100 14030725 02 08 2019 EN DE DE 12 Technische Daten 170 6 69 448 17 64 330 12 99 390 15 35 Abmessungen in mm in Ansicht von oben Ansicht von der Seite links Ansicht von hinten...

Page 42: ...13 Zubeh r Zubeh r Bestellcode Beschreibung CTX A K9 Transportkoffer TB Netzkabel f r die EU EU f r die Schweiz CH f r UK UK f r USA Kanada US H lsenwechselwerkzeug RT Bestellangaben f r Ihre Anfrage...

Page 43: ...43 WIKA operating instructions model CTD9100 1100 14030725 02 08 2019 EN DE...

Page 44: ...Alexander Wiegand SE Co KG Alexander Wiegand Stra e 30 63911 Klingenberg Germany Tel 49 9372 132 0 Fax 49 9372 132 406 info wika de www wika de 44 WIKA subsidiaries worldwide can be found online at w...

Reviews: