DistilLED water
Heating up:
Period
Cooling down:
Period
2 °C to 25 °C
5:31 min
90 °C to 75 °C
3:09 min
25 °C to 50 °C
6:49 min
75 °C to 50 °C
7:06 min
50 °C to 75 °C
8:07 min
50 °C to 25 °C
10:18 min
75 °C to 90 °C
6:19 min
25 °C to 2 °C
14:52 min
Silicone oil 5 CS
Heating up:
Period
Cooling down:
Period
-30 °C to -25 °C
0:56 min
120 °C to 100 °C
32:24 min
-25 °C to -15 °C
1:06 min
100 °C to 75 °C
3:40 min
-15 °C to 0 °C
1:18 min
75 °C to 50 °C
4:48 min
0 °C to 25 °C
2:46 min
50 °C to 25 °C
6:41 min
25 °C to 50 °C
2:37 min
25 °C to 0 °C
8:50 min
50 °C to 75 °C
3:10 min
0 °C to -15 °C
10:36 min
75 °C to 100 °C
4:23 min
-15 °C to -25 °C
15:01 min
100 °C to 120 °C
5:05 min
-25 °C to -30 °C
23:19 min
GB
Temperature Dry Well Calibrators/Micro Calibration Baths
CTD9100/CTB9100
11263911 04/2007 GB/D
WIKA Operating Instructions Temperature Dry Well Calibrators/Micro Calibration Baths
1.4 Heating and cooling periods
1.4.1 CTB9100-165
Measuring conditions:
·All heating and cooling periods are related to a
reference Pt100 probe, Ø 6 mm, stainless steel
sheath
·The reference sensor is located 5 mm above the
protective basket in the middle of the tank
·All heating and cooling periods do not include
stabilisation time
·The measurements were carried out at a room
temperature of approx. 23 °C with bath without
cover