Technical data CTD9100
Display range/Setting range
CTD9100-165
-50.0 °C to +165.0 °C in 0.1 °C resolution
CTD9100-450
0.0 °C to +450.0 °C in 0.1 °C resolution
CTD9100-650
0.0 °C to +650.0 °C in 0.1 °C resolution
Temperature setting range
CTD9100-165
between -30 °C and 165 °C
CTD9100-450
between room temperature and 450 °C
CTD9100-650
between room temperature and 650 °C
Block temperature control
via PID controller
Setting the temperature
via the P key and the
5
or
key to 0.1 °C
fine adjustment occurs automatically
Accuracy
CTD9100-165
+/- 0.15 K ... 0.25 K
CTD9100-450
+/- 0.30 K ... 0.50 K
CTD9100-650
+/- 0.30 K ... 0.80 K
Control stability
+/- 0.05 °C
Block temperature display
4-digit, 7-segment LED, 7 mm high
red = PV, green = SV
Display resolution
+/- 0.1 °C
Ambient temp. influence
(0...50 °C) to the accuracy
+/- 0.02 °C/°C
Excess temperature behaviour
temperature fuses interrupt the power supply if there is excess
temperature inside the housing
Sensor break behaviour
the control is switched off
Display unit
°C or °F (optional)
Display for sensor break
- - - -
Detection speed
130 ms
Controller outputs
1 x voltage output for control of the solid state relay (8 mA/8 VDC),
heater control
1 x voltage output for control of the solid state relay (8 mA/8 VDC),
cooling control
1 x relay SPDT (8 A-AC1, 3 A-AC3/250 VAC)
100,000 switching cycle, fan control
Power supply
CTD9100-165
90…240 VAC +/- 10 %, 50/60 Hz
CTD9100-450
230 VAC -15 %/+10 %, 50/60 Hz
CTD9100-650
230 VAC -15 %/+10 %, 50/60 Hz
optional 115 VAC -20 %/+10 %, 50/60 Hz
GB
Temperature Dry Well Calibrators/Micro Calibration Baths
CTD9100/CTB9100
11263911 04/2007 GB/D
WIKA Operating Instructions Temperature Dry Well Calibrators/Micro Calibration Baths
0
1. Technical data CTD9100