GB/D/F/E 01/2009
1
WIKA Operating instructions/Betriebsanleitung/Mode d'emploi/Instrucciones de servicio AC-1
GB/D/F/E 01/2009
1
WIKA Operating instructions/Betriebsanleitung/Mode d'emploi/Instrucciones de servicio AC-1
Failure
Possible cause
Procedure
Output signal unchanged after change
in pressure
Mechanical overload through over-
pressure
Replace instrument; if failure reoccurs,
consult the manufacturer *)
Wrong supply voltage or current spike Replace instrument
No/False output signal
Incorrectly wired (e.g. Connected as
2-wire instead of 3-wire system)
Follow pin assignment (see Instrument
Label / Operating Instructions)
No output signal
No/incorrect voltage supply or current
spike
Adjust the voltage supply to correspond
with the Operating Instructions *)
Cable break
Check connections and cable
Signal span too small
Power supply too high/too low
Correct the power supply in line with
the Operating Instructions
Signal span too small
Mechanical overload through over-
pressure
Replace instrument; if failure reoccurs,
consult the manufacturer *)
Signal span drops off
Moisture present (e.g. at the cable tail) Install the cable correctly
Signal span erratic
Electromagnetic interference source in
the vicinity, e.g. inverter drive
Shield the device; shield the cables;
remove the interference source.
Instrument not grounded
Ground instrument
Strong fluctuations in the power supply Stabilise the power supply; smooth it
(i.e.; remove interferences)
Violent fluctuations in the process
media pressure
Damping; consult with manufacturer
9. Trouble shooting
GB
In case of unjustified reclamation we charge the reclamation handling expenses.
*) Make sure that after the assembly the unit is working properly. In case the error continues to exist send in the instrument for
reparation (or replace the unit).
Please verify in advance if pressure is being applied (valves / ball valve etc. open) and if the
right voltage supply and the right type of wiring (2-wire / 3-wire) has been chosen?
If the problem persists, contact our sales department.
USA, Canada
If the problem continues, contact WIKA or an authorized agent for assistance. If the pres-
sure transmitter must be returned obtain an RMA (return material authorization) number and
shipping instructions from the place of purchase. Be sure to include detailed information about
the problem. Pressure transmitters received by WIKA without a valid RMA number will not be
accepted.
Process material certificate (Contamination declaration for returned goods)
Purge / clean dismounted instruments before returning them in order to protect our employees
and the environment from any hazard caused by adherent remaining media.
Service of instruments can only take place safely when a Product Return Form has been
submitted and fully filled-in. This Return Form contains information on all materials with which
the instrument has come into contact, either through installation, test purposes, or cleaning.
You can find the Product Return Form on our internet site (www.wika.de / www.wika.com).
9. Trouble shooting / 10. Storage, disposal
GB
10. Storage, disposal
When storing or disposing of the pressure transmitter, take precautions with
regard to remaining media in removed pressure transmitters. We recom-
mend cleaning the transmitter properly and carefully. Remaining media in
the pressure port may be hazardous or toxic!
Dispose of instrument components and packaging materials in accordance with the
respective waste treatment and disposal regulations of the region or country to which
the instrument is supplied.
Disposal
WIKA reserves the right to alter these technical specifications.
!
Warning