EN
30
40202755.03 09/2017 EN/DE
WIKA operating instructions model A2G-25
3. Select K factor: Between 0.001 … 9999.000
If “MANUFACTURER” is selected in step 1, then only enter the
manufacturer and ventilator model. The K factor is automatically
determined by the measuring instrument.
If a measuring probe or a different ventilator is used, then the K factor
must be entered.
Move the “DOWN” button once, shortly.
▶
“K-VALUE” menu item is displayed
Move the “SELECT” button once, shortly, in order to
activate the “K-VALUE” selection.
▶
Selection flashes
Use “UP” or “DOWN” to enter the desired numbers.
Move the “SELECT” button to the left and right to
select the decimal point.
▶
The “K-VALUE” is displayed
Move the “SELECT” button once, shortly, in order to
accept the selection.
K-VALUE
1,100
K-VALUE
1,100
K-VALUE
0061.130
K-VALUE
0061.130
6. Menu navigation