EN
24
40202755.03 09/2017 EN/DE
WIKA operating instructions model A2G-25
5. Select the maximum differential pressure for the optimisation of
the output accuracy
Move the “DOWN” button once, shortly.
▶
“P OUTPUT MAX” menu item is displayed
Move the “SELECT” button once, shortly, in order to
activate the differential pressure selection.
▶
Selection flashes
Use “UP” or “DOWN” to find the desired unit.
▶
Selection is displayed
Move the “SELECT” button once, shortly, in order to
accept the selection.
Max. differential pressures in connection with the selected measuring
instrument variant.
Variant
Pressure unit
Pa
kPa
mbar
mmWC
inWC
A2G-25-1000 Pa
100 ... 1,000
0.1 ... 1.0
1.0 ... 10
10 ... 100
0.4 ... 4.0
A2G-25-2000 Pa
200 ... 2,000
0.2 ... 2.0
2.0 ... 20
20 ... 200
0.8 ... 8.0
A2G-25-5000 Pa
500 ... 5,000
0.5 ... 5.0
5.0 ... 50
50 ... 500
2.0 ... 20
A2G-25-7000 Pa
700 ... 7,000
0.7 ... 7.0
7.0 ... 70
70 ... 700
2.5 ... 30
P OUTPUT MAX
20.00 mbar
P OUTPUT MAX
20.00 mbar
P OUTPUT MAX
18.50 mbar
P OUTPUT MAX
18.50 mbar