Power supply gesis RM-PS 12/5
46
Wieland Electric | BA000580 | 6/2009 (Rev. A)
10.5
Technical Data
Mains power input
(connection X1)
85…264
V AC
(for 85
V AC 80
% power, from 115
V AC 100%)
120…370
V DC
Frequency range
47…440
Hz
Efficiency
typically 73
%
AC current consumption
typically 0.12
A/115
V AC
typically 0.08
A/230
V AC
Output (connection X2)
Rated voltage
12,5
V DC ±
4
% (SELV)
(I/O isolation according to EN 60601-1)
Rated current
400
mA
Power
5
W
Short circuit withstand capability Yes
Overload protection
Yes
Overtemperature protection
Yes
Connection type (X1 and X2)
Screw clamp terminals
0.14 to 4
mm², single-core
0.14 to 2.5
mm², finely stranded
6.5
mm stripped in each case
Electrical safety
Protection class
none
Degree of protection
IP 20
Degree of soiling
2
Surge voltage category
III
Rated insulation voltage
250 V external conductor to N or PE
Electrical isolation
Air gap/creepage paths > 5.5
mm (supply/output)
Environmental conditions
Ambient operating temperature
−
5°C to +45°C
Storage temperature
−
25°C to +70°C
relative humidity
5% - 93%
Moisture condensation
not permitted
General data
Housing material
Plastic, halogen-free
Housing colour
black
Behaviour in fire
V2 according to UL (housing)
Weight
approx. 110
g
Dimensions
see "Dimensions, Connections and Function Elements"
Height including TH 35-7.5
mounting rail
52
mm
CE marking
according to EMC Directive (residential and functional
buildings), Low Voltage Directive
Mounting
on TH
35 mounting rail