Instructions for Operation and Care of Water Bath SP-1600-A

SET-UP INSTRUCTIONS

Operator trainees need to:

Ÿ

read and understand this manual.

Ÿ

be trained in occupational therapy protocols.

Training

Skills:

Operators  using  the  Water  Bath  need  a  working 
knowledge of occupational therapy procedures.

WARNING

!

Untrained operators can cause injury or be injured. 

Only trained personnel should operate the Water Bath.

WARNING

!

OPERATOR SKILLS AND TRAINING

INSPECTING THE WATER BATH™

Ÿ

Is the outlet equipped with a functioning GFI?

Ÿ

Does the drain valve open and close properly?

Ÿ

Is  a  properly-grounded  and  voltage-matched 

receptacle available?

Inspection Checklist

Ÿ

Is the Water Bath free of excessive wear?                                     

 Inspect the contents for damage.

If there is visible damage to the instrument upon receipt, 
inform the shipping company and Whitehall immediately.

This Whitehall product has been carefully packaged at 
the factory to minimize the possibility of damage during 
shipping.

 Inspect the packaging for external signs of     

damage.

WARNING

!

Do not attempt to operate this equipment if there is 
evidence of shipping damage or you suspect the unit 
is damaged. Damaged equipment may present 
additional hazards to you. Contact Whitehall technical 
support for advice before attempting to plug in and 
operate damaged equipment. 

WARNING

!

2. Unpacking your Water Bath

Remove  all  shipping  &  packaging  material.  Remove  all 
items  from  inside  of  tank  such  as  Power  Cordset, 
Warranty  Registration  Card,  and  Operation  & 
Maintenance Manual.

(1) Power Cordset 

1. Your new Water Bath includes the following:

(1) Heating Appliance

(1) Operation & Maintenance Manual

NOTE: If you are missing any of the above, 
contact WHITEHALL MANUFACTURING
toll free telephone no. (800) 782-7706 or (626) 968-6681

Place  the  Water  Bath  at  a  convenient  height  &  location 
such as on a table or scratch-resistant counter top. 

IMPORTANT:

  Appliance  must  be  placed  on  a  level 

surface! Water Bath must be positioned so to allow ease 
of disconnecting Power Cordset if necessary.

3. Location of your Water Bath

Summary of Contents for SP-1600-A

Page 1: ...BATH INSTRUCTIONS FOR OPERATION AND CARE OF 800 782 7706 626 968 6681 www whitehallmfg com P O Box 3527 City of Industry CA 91744 0527 U S A Division of Acorn Engineering Company 6900 117 000_REV_B Revised May 31 2019 ...

Page 2: ...personal injury or damage to the equipment is possible WARNING To avoid electric shock connect the instrument to properly earth grounded GFCI protected 3 prong receptacles only Failure to observe this precaution can result in severe injury Ÿ ALWAYS allow the appliance to cool down before draining water or moving unit Ÿ DO NOTuse the lid handle to lift or transport appliance Ÿ NEVER leave appliance...

Page 3: ...3 Preparing to fill your Water Bath 4 Filling your Water Bath 4 Preparing to turn ON your Water Bath 4 Turning ON your Water Bath 4 Changing the temperature on your Water Bath 4 Using Your Water Bath 5 Table of Contents Emptying Your Water Bath 5 Care Cleaning of Your Water Bath 5 Product Specifications Model SP 1600 A 2 Table of Contents 1 FAQ Troubleshooting 6 7 Troubleshooting for Temperature C...

Page 4: ...ial heat up temperature duration is based on the splint material manufacturer s requirements and instructions The Water Bath is primarily constructed of Type 304 Stainless Steel material is internally insulated and has variable temperature settings with a LED display and a concealed wire wound heating element The Water Bath is a water tank capable of heating water within a set temperature range of...

Page 5: ...mmediately This Whitehall product has been carefully packaged at the factory to minimize the possibility of damage during shipping Inspect the packaging for external signs of damage WARNING Do not attempt to operate this equipment if there is evidence of shipping damage or you suspect the unit is damaged Damaged equipment may present additional hazards to you Contact Whitehall technical support fo...

Page 6: ...pped to lower temperature Your water temperature is adjustable and can be set to reach splint materials recommended operating temperature See splint manufacturer s instructions Note Most splint material on the market today has an operating range between 113 F 45 C to 200 F 93 C Your Water Bath is capable of approximate maximum temperature of 200 F 93 C depending on amount of water in unit The ther...

Page 7: ... the open position and will not close place index finger on the left side lid stay reset button and work lid up and down until lid stays synchronize and allow lid to close USING YOUR WATER BATH EMPTYING YOUR WATER BATH The appliance should be routinely cleaned to help ensure the appearance as well as years of reliable service Routine cleaning helps to remove contaminants from the stainless steel s...

Page 8: ...ck disconnect terminals wire connectors wiring etc may have become faulty A qualified electrical technician should perform an electrical continuity inspection of the entire electrical system using a Multi Meter to identify possibly bad electrical component s Replace electrical component s as required My Water Bath temperature raises and lowers by itself Equipment Circuit Protection fuse s may have...

Page 9: ...t Replace the thermostat The thermostat is set too high The thermostat is broken Splint material too hot The ON OFF switch is broken Replace the switch Turn on the circuit breaker Replace the thermostat Replace the heating element Circuit breaker is off Thermostat is not working Heating element is burned out is ON plugged in and switch even though unit is Water doesn t get hot Splint material too ...

Page 10: ... in the above figure identifies the location of the bonding terminal which is bonded to conductive accessible parts of the enclosure for safety purposes Protective Conductor Terminal Alternating Voltage or Current Fuse Location Symbols on labels Use only Time Delay Fuse See Parts List Page 10 Item 26 FUSE 1 FUSE 2 WHT BLK YEL GRN WHT BLK DPST SWITCH HEATING ELEMENT OVER TEMP SENSOR OPERATING THERM...

Page 11: ... CA 91744 0527 U S A Phone 800 782 7706 626 968 6681 Fax 626 855 4862 Web www whitehallmfg com 9 Instructions for Operation and Care of Water Bath SP 1600 A 15 5 3 6 7 8 1 4 21 17 13 12 10 14 16 9 18 22 28 27 29 11 2 30 19 18 24 23 26 25 Fuse 1 Fuse 2 20 ...

Page 12: ...er Control Knob 9 6903 152 000 1 4 NPT SST Ball Valve 10 6903 160 000 1 4 NPT x 1 4 Tube Drain Barb SST 11 5903 715 199 Bottom Cover Assembly 12 7000 415 000 Self Retaining J Nut 13 6502 100 000 8 32 x 3 8 Phil Pan Head Screw 14 6900 138 000 Pull Handle 4 5903 741 001 Side Hinge Bracket L Assembly 15 6900 154 000 10 24 x 1 2 Pan Head Screw SST 23 6903 150 000 Operating Thermostat 24 6903 556 000 H...

Page 13: ... A Phone 800 782 7706 626 968 6681 Fax 626 855 4862 Web www whitehallmfg com 11 Instructions for Operation and Care of Water Bath SP 1600 A Switch Heating Element Thermostat Digital Thermometer Power Inlet Fuse Holder Fuse Operating Thermostat ELECTRICAL PARTS DISASSEMBLY ...

Page 14: ...h have been damaged by other causes such as mishandling or improper care or abnormal use The repair or replacement of the defective materials shall constitute the sole remedy of the Buyer and the sole remedy of Whitehall under this warranty Whitehall shall not be liable under any circumstances for incidental consequential or direct charges caused by defects in the materials or any delay in the rep...