background image

WHITEHALL MANUFACTURING

 • P.O. BOX 3527 • City of Industry, CA 91744-0527 U.S.A

Phone (800) 782-7706 • (626) 968-6681 • Fax (626) 855-4862 • Web: www.whitehallmfg.com

15

Instructions for Operation and Care of Dickson Paraffin Baths

DICKSON PARAFFIN BATH 
MODEL PB-105

  

FUSE 1

FUSE 2

BLK

WHT

YEL/GRN

WHT

WHT

GND

POWER

INLET

SK

RESISTOR

GREEN

LIGHT

RED

LIGHT

BLK

L

N

HEATING

ELEMENT

TIMER

CONTROL

BLK

BLK

BLK

OPERATING

THERMOSTAT

LOW TEMP

THERMOSTAT

HIGH TEMP

THERMOSTAT

FUSE 1

FUSE 2

BRN

BLU

YEL/GRN

BLU

BLU

GND

POWER

INLET

SK

RESISTOR

GREEN

LIGHT

RED

LIGHT

BRN

L

N

HEATING

ELEMENT

TIMER

CONTROL

BRN

BRN

BRN

OPERATING

THERMOSTAT

LOW TEMP

THERMOSTAT

HIGH TEMP

THERMOSTAT

DICKSON PARAFFIN BATH 
MODEL PB-105B

  

Summary of Contents for Dickson PB-104

Page 1: ...fg com 6900 189 000 INSTRUCTIONS FOR OPERATION AND CARE OF INSTITUTIONAL PARAFFIN BATH P O Box 3527 City of Industry CA 91744 0527 U S A Revised October 2018 PB 104 PB 104B PB 105 PB 105B Division of...

Page 2: ...nouvel appareil avant de l utiliser l EFFECTUER LA MISE LA TERRE de votre appareil DE FA ON APPROPRI E Brancher l appareil dans une prise disjoncteur diff rentiel de fuite la terre DDFT ce qui offre...

Page 3: ...Fuse 6 Wiring Diagram for PB 104 14 Wiring Diagram for PB 105 15 Warranty Info 16 Cleaning the Bath 6 Parts Diagram for PB 105 12 Wiring Diagram for PB 107 13 Paraffin Refills 6 Directions for Treatme...

Page 4: ...e the tank from the atmosphere and thereby minimize heat conduction and do contribute to the stability of the bath temperature These moldings in slightly extending over the inside of the tank also ser...

Page 5: ...ensions 14 1 2 L x 7 W x 10 D 368 L x 178 W x 254 D Tank Capacity 18 lbs 8 16 kgs Electrical Requirements Inlet Voltage 100 120 V Frequency 50 60 Hz 3 8 A Inlet Voltage 200 240 V Frequency 50 60 Hz 1...

Page 6: ...4 Instructions for Operation and Care of Dickson Paraffin Baths DICKSON PARAFFIN BATH MODELS PB 105 PB 105B High Heat Sanitizing Indicator Light Low Heat Treatment Temperature Indicator Light High Hea...

Page 7: ...red 10 to 15 minutes for hands 15 to 30 minutes for arms Because of the high heat input of the paraffin the bath must be used judiciously If circulatory and sensory changes are present the application...

Page 8: ...Bath should be disconnected from the electrical outlet and the drain ring removed The sediment and sludge may then be drained the bottom flushed and the slats cleaned PARAFFIN REFILLS Whitehall Manufa...

Page 9: ...the wax up incrementally by small amounts 10 Old wax not melting properly and hardening over time after continual reuse There is a short circuit in the unit The unit must be opened up for repairs 9 Wa...

Page 10: ...ing x x 15 0710 216 001 Fuse 5 x 20mm 10A Fast Acting x x 16 0710 230 001 Fuse Holder 5 x 20mm x x x x x x 17 6900 130 000 Knob For Timer or Switch x x x x x x 18 6900 135 000 Light Rectangle Green 12...

Page 11: ...stor Assembly 12K 220V x 34 6902 470 001 Resistor Assembly 3K 120V x 35 6902 471 001 Resistor Assembly 12K 220V x 36 6902 374 000 Slat Plexiglass x x 37 6902 424 000 Slat Plexiglass x x 38 6505 546 00...

Page 12: ...A 91744 0527 U S A Phone 800 782 7706 626 968 6681 Fax 626 855 4862 Web www whitehallmfg com 10 Instructions for Operation and Care of Dickson Paraffin Baths DICKSON PARAFFIN BATH MODEL PB 107 24 1 44...

Page 13: ...7 U S A Phone 800 782 7706 626 968 6681 Fax 626 855 4862 Web www whitehallmfg com 11 Instructions for Operation and Care of Dickson Paraffin Baths DICKSON PARAFFIN BATH MODEL PB 104 47 33 20 18 28 6 4...

Page 14: ...U S A Phone 800 782 7706 626 968 6681 Fax 626 855 4862 Web www whitehallmfg com 12 Instructions for Operation and Care of Dickson Paraffin Baths DICKSON PARAFFIN BATH MODEL PB 105 45 18 20 48 17 19 21...

Page 15: ...lmfg com 13 Instructions for Operation and Care of Dickson Paraffin Baths FUSE FUSE BLU BRN YEL GRN BLU BRN GND POWER INLET L N HEATING ELEMENT THERMOSTAT RED LIGHT TIMER CONTROL GREEN LIGHT L200 TWIS...

Page 16: ...B 104 FUSE 1 FUSE 2 BLK WHT YEL GRN WHT WHT GND POWER INLET SK RESISTOR GREEN LIGHT RED LIGHT BLK L N HEATING ELEMENT TIMER CONTROL BLK BLK BLK OPERATING THERMOSTAT LOW TEMP THERMOSTAT HIGH TEMP THERM...

Page 17: ...B 105 FUSE 1 FUSE 2 BLK WHT YEL GRN WHT WHT GND POWER INLET SK RESISTOR GREEN LIGHT RED LIGHT BLK L N HEATING ELEMENT TIMER CONTROL BLK BLK BLK OPERATING THERMOSTAT LOW TEMP THERMOSTAT HIGH TEMP THERM...

Page 18: ...r direct charges caused by defects in the materials or any delay in the repair or replacement thereof This warranty is in lieu of all other warranties expressed or implied Product maintenance instruct...