PPG Heart Rate Simulator Test System | User Manual
3.2.2
Fine Tune Mode
Below are the details of Fine Tune Mode interface:
01
–
Synchronized Pulse
:
Select whether HRS LED flashes synchronously with DUT LED.
02
–
Waveform Setting
:
Allows user to select Sine wave, Triangle wave, and PPG wave in the test.
03
–
DC Value
:
This parameter simulates the reflection of different brightness levels.
Range: 100 mv to 2500 mV.
04
–
AC Value
:
This parameter simulates AC changes. Range: 0.75 mv to 25 mV.
05
–
Noise Frequency
:
Allows user to select the noise in 50Hz, 60Hz or 1000Hz (1kHz)
frequency.
06
–
Noise Amplitude
:
This parameter simulates different amplitudes of noise. Peak-to-peak
voltage range: 0.01mVpp to 2mVpp.
07
–
BPM Value
:
Adjust this parameter to change Heart Rate setting. Range: 30 to 300
Beats-Per-Minute.
08
–
Display Settings
:
Automatically calculate the output PI value of HRS.