16
PPG Heart Rate Simulator Test System | User Manual
Former
Version
-
Background color is all white
New
Version
-
Background color of PD Signal Display section is gray
-
Two different PD scale ranges: “1000” and “10000”
1.5 Cautions
➢
We suggest warm up HRS unit for at least 5 minutes before testing.
➢
Before using HRS unit, please assure the distance between HRS diffusion
film and DUT PPG sensor is fixed. Also, it requires to control the ambient
light as it would influence the test result as well.
➢
PD sample rate of HRS is lower than 8ksps (Normal) and 30ksps (High
Resolution)
➢
If HRS software shows up “
S
aturated” in PD Signal Display section, it
means the brightness of DUT is over the detection range of HRS PD.