22
PPG Heart Rate Simulator Test System | User Manual
04
–
PI Value
:
This parameter simulates different AC/DC ratio. Please refer to
Section1.1.2 for more details.
05
–
Save to HRS
:
Save DC Level and PI value into HRS internal system memory as the
startup default values.
06
–
BPM Value
:
Adjust this parameter to change Heart Rate setting. Range: 30 to 300
Beats-Per-Minute.
07
–
Save as Mode
:
This feature allows users to save their own settings into HRS internal
memory and be later used in HRS standalone mode by pressing the mode
button. There are 3 Modes can be saved (Mode A/B/C), and each Mode
contains 3 parameters (DC Level, PI value, and BPM value).
PD Relative Settings
:
(refers to Section 1.1.4)
08
–
Filter Settings:
Click this b
utton to set “Pass/Fail” condition for
sorting LED. Please refer
to Section 3.2.1.1 for details.
09
–
Start/Stop Filter:
After clicking this button, Start or Stop to judge whether the peak value of
LED is in range. Please refer to Section 3.2.1.1 for details.
10
–
Filter Result:
Show the result or remaining seconds. Please refer to Section 3.2.1.1 for
details.
11
–
Freezed:
Select whether stay “Display PD Sampling” and “Max Peak” parts no
change.
12
–
Max Peak:
Show the maximum peak value during the specific time period.
13
–
PD range Settings:
C
hange the PD scale to “1000” or “10000”. Select the scale to “1000”
while testing DUTs with lower LED brightness. Select the scale to “10000”
while testing DUTs with higher LED brightness.
14
–
Time Scale Setting:
Select and set the display time interval of PD Sampling.
15
–
Display PD Sampling:
Show the received data from HRS PD. It means the transmitted data from
DUT LED.