DFS360
|
User Manual
8
2.3
Energy Measurement Specifications
Table 3: Energy Measurement Specifications
2.4
ECG Signal Specifications
Table 4: ECG Signal Specifications
Items
Specifications
Ventricular Fibrillation -
Coarse
V
p-p
= 3 mV
Ventricular Fibrillation -
Fine
V
p-p
= 0.5 mV
Ventricular Tachycardia
(
≒
180BPM)
V
p-p
= 3 mV
Ventricular Tachycardia
Plus (
≒
234 BPM)
V
p-p
= 3 mV
Atrial Fibrillation
V
p-p
= 1.2 mV
Items
Specifications
Maximum Energy
Range: Up to 600 J
Accuracy:
±
1 % of reading
±
1 J for up to 360J;
±
1 % of reading
±
1 J, typical for 360J ~ 600J
Voltage
Range: Up to 3000 V
Accuracy:
±
(1% of r 1V) for 450V ~ 3000V & -450V ~ -3000V;
± (1% of r 2V) for -450V ~ 450V
Load Resistance
Resistance: 50
Ω
Pulse Width
Range: 1.0 ~ 48.0 ms
Accuracy: ±0.05 ms
Sample rate
250 kHz (4 μs sample)
Maximum Average
Power
DFS360: 12 W, for 1 defib. Pulse of 360 J every 30 sec.
DFS360D: 24 W, for 1 defib. pulses of 360 J every 15 sec.