WHALETEQ DFS360 User Manual Download Page 25

 

 

 

 

 

DFS360

 

|

 

User Manual 

24 

 

5.1.2

 

Customize ECG Waveform (optional function) 

 

After purchasing this function, users will receive an Activation Key. Input the 
key to activate the function and bring up the ECG Waveform window. 
Users can add 10 sets of user-defined waveforms or raw data at most. 

 

Figure 20: License Activation

 

 
 
 

Figure 21: ECG Waveform

 

 

 

Table 7: Customize ECG Waveform Function Introduction 

Item 

Name 

Description 

Add 

Add user-defined waveforms or raw data. 

Summary of Contents for DFS360

Page 1: ...DFS360 User Manual WHALETEQ DFS360 DFS360D User Manual PC Software Version 1 0 1 18 Revision 2022 08 08 ...

Page 2: ...d the programs as is without warranty of any kind either expressed or implied including but not limited to the implied warranties of merchantability or fitness for a particular purpose This document could contain technical inaccuracies or typographical errors Changes are periodically made to the information herein these changes will be incorporated in future revisions of this document WhaleTeq Co ...

Page 3: ... 12 4 Getting Started 13 4 1 PC System Requirements 13 4 2 Software Installation 14 4 3 USB Driver Installation 14 4 4 Microsoft Net Framework Installation 15 4 5 PC Software and Firmware Update 16 5 PC Software Operation 16 5 1 Software Function Introduction 16 5 1 1 Function Buttons Operation 19 5 1 2 Customize ECG Waveform optional function 24 5 2 PC Software Operation Instructions 28 5 3 Over ...

Page 4: ...l 3 5 4 1 Command Function 34 6 Standalone Operation 35 7 Calibration and Validation 35 8 Troubleshooting 36 9 Cautions 36 10 Ordering Information 37 11 Package Contents 40 12 Revision History 41 13 Contact WhaleTeq 41 ...

Page 5: ...ustomize ECG Waveform Function Introduction 24 Table 8 Shock Result Introduction 30 Table 9 Command Function 34 Table 10 Ordering info for DFS360 37 Table 11 Ordering info for DFS360D 37 Table 12 Optional DFS360 DFS360D Paddle Box Set 38 Table 13 Optional Software Add On Pack 39 Table 14 Optional Tool 39 Table 15 Optional Calibration Service and Warranty Extension 39 Table 16 Package Contents for ...

Page 6: ...re 13 Set the Time Range and Display of Voltage or Current 20 Figure 14 Error Report Window 21 Figure 15 Set a Designated Range of Time 21 Figure 16 Check Desired Test Results for Analysis 22 Figure 17 Consistency and Availability Analysis Top The Consistency of Shock Waveforms of Multiple DUTs Bottom The Availability of a Single DUT 23 Figure 18 Standard Waveforms 23 Figure 19 Adjustable BPM Freq...

Page 7: ...ng the characteristic points of each kind of waveform DFS360 DFS360D is also capable of displaying the real defibrillation waveforms These features can be consolidated into the computer software of AED or defibrillator manufacturers What s more DFS360D is designed to withstand more intense defibrillation shocks over longer stretches it takes shocks at 15 second intervals meaning a boost of testing...

Page 8: ...us Power Communications USB Type C connector Operation Modes Remote PC Software via API and SDK Mechanical Housing ABS Plastic Size L x W x H 19 0 x 12 5 x 6 0 cm Weight DFS360 665g DFS360D 815g 2 2 Compliance Specifications Table 2 Compliance Specifications Items Specifications Safety Standards CE IEC EN61010 1 2010 A1 2016 Pollution degree 2 61010 2 030 2017 Electromagnetic Compatibility Standar...

Page 9: ... Plus 234 BPM Vp p 3 mV Atrial Fibrillation Vp p 1 2 mV Items Specifications Maximum Energy Range Up to 600 J Accuracy 1 of reading 1 J for up to 360J 1 of reading 1 J typical for 360J 600J Voltage Range Up to 3000 V Accuracy 1 of reading 1V for 450V 3000V 450V 3000V 1 of reading 2V for 450V 450V Load Resistance Resistance 50Ω Pulse Width Range 1 0 48 0 ms Accuracy 0 05 ms Sample rate 250 kHz 4 μs...

Page 10: ...rn Max 5K Sample Rate Up to 10 sets with 100 seconds max each set Normal Sinus Rhythm NSR 30 to 240 increment by 1 BPM Amplitude range Vp p 0 5 5 0mV step 0 5 mV 3 Instrument Familiarization 3 1 Top Figure 1 Top Side of DFS360 DFS360D 3 2 Front Figure 2 Front Side of DFS360 DFS360D A ...

Page 11: ...USB Port Use to supply power Also it can be used to connect to the computer for sending commands and transferring data C Status LED Indicate the working state of DFS360 DFS360D Mode Normal Heat Alert Standalone Operation Green light flashes every 2 seconds Red light flashes every 2 seconds AP Connected ECG Playing AP Connected ECG NOT Playing Green light always on Red light always on B C ...

Page 12: ...0 Bottom DFS360D Figure 5 Right Side of DFS360 DFS360D D Air Inlet Allow ventilation during long hours of operation E External Load Box Port Connect this port to the external load box with various resistance values for tests F Air Outlet Allow ventilation during long hours of operation D F E D ...

Page 13: ...o not put the AED cable on DFS360 DFS360D main console The AED cable should not twist together Figure 7 Correct Wiring Diagram of the AED Cable 2 After connecting the AED cable connect DFS360 DFS360D and PC Laptop with the USB cable The USB cable and PC Laptop should not be placed near the AED cable The USB cable should not twist with the AED cable ...

Page 14: ...60D via USB module PC requirements Windows PC Windows 7 or later suggest to use the genuine version Microsoft NET 4 0 or higher Administrator access essential for installing software driver and Microsoft Net Framework 1 5 GHz CPU or higher 1GB RAM or higher1 1 Relative to normal PC processing there is no special use of PC speed However there has been noted a slow increase in system RAM usage over ...

Page 15: ...work 4 0 are both installed 4 3 USB Driver Installation If Windows Device Manager can t recognize WhaleTeq s product please follow the instructions below to install Microchip USB Driver For Microsoft Windows 10 As Windows 10 has built in Microchip USB Driver there is no need to install any drivers It just takes a while for Windows Device Manager to recognize and install the driver For Microsoft Wi...

Page 16: ...inf from WhaleTeq s website This driver is provided by Microchip and is intended to be used with PIC microprocessors which have built in USB functions When the USB module is connected for the first time select Manual Installation and point to the folder containing the above file Then continue to follow the instructions to finish the installation There may be a warning that the driver is not recogn...

Page 17: ...ware please contact WhaleTeq s service team service whaleteq com 5 PC Software Operation 5 1 Software Function Introduction When opening DFS360 AP User will see the figure below on the screen Figure 9 AP Interface Table 5 AP Interface Function Introduction Item Name Description A Add Add one or more ECG waveform s to the Test List B A F L G E I H K C D J ...

Page 18: ... Joules parameters and characteristic points of the shock K Settings Check for updates set the customized ECG waveforms choose naming methods of test results and select the language etc L Function Buttons Top to bottom Folder Open the folder storing test results in the local computer Waveform Details Display positions of the parameters and characteristic points on the defibrillation waveform Graph...

Page 19: ...mode Only NSR with adjustable BPM frequency and amplitude Ventricular Fibrillation Coarse Ventricular Fibrillation Fine Ventricular Tachycardia Ventricular Tachycardia Plus the faster VT Atrial Fibrillation Asystole NSR Right click the Settings button and the settings below will show Figure 10 Setting List Table 6 Setting List Introduction Item Name Description M Check for Updates Check whether th...

Page 20: ...tivate this function P Language Switch the language of the DFS360 AP Note The languages include Traditional Chinese Simplified Chinese and English 5 1 1 Function Buttons Operation Function Buttons on the right of DFS360 DFS360D AP screen will be elaborated from top to bottom 1 Folder After every test the system will automatically save the test result in three documents including the document of ra...

Page 21: ...to display the positions of parameters and characteristic points on the defibrillation waveform Figure 12 Waveform Details 3 Graph Setup Users can set the time range 30ms 40ms or 60ms and display of voltage or current for the Shock Waveform Figure 13 Set the Time Range and Display of Voltage or Current ...

Page 22: ...range of test results to superimpose multiple test waveforms to compare for confirming the consistency of quality on multiple DUTs Besides the mean value and standard deviation are automatically calculated as a reference for checking the availability of a single DUT 1 Click Consistency and Availability Analysis button to bring up the Analysis window Users can click the Filter button to set a desig...

Page 23: ... frame on the lower section to select a designated range of test results to superimpose their shock waveforms on the upper section At the upper right corner on the lower section the green circle on the figure below the auto calculated mean value and standard deviation show as a reference for checking whether the joules of each test result is standard By superimposing waveforms users can confirm th...

Page 24: ...k Waveforms of Multiple DUTs Bottom The Availability of a Single DUT 6 Standard Waveforms Select Sine Square or Triangle waveforms with adjustable BPM frequency and amplitude for essential performance test and calibration Figure 18 Standard Waveforms Figure 19 Adjustable BPM Frequency and Amplitude ...

Page 25: ...ion Key Input the key to activate the function and bring up the ECG Waveform window Users can add 10 sets of user defined waveforms or raw data at most Figure 20 License Activation Figure 21 ECG Waveform Table 7 Customize ECG Waveform Function Introduction Item Name Description Q Add Add user defined waveforms or raw data R S T Q ...

Page 26: ... more than 100 seconds it will be automatically cut off when loaded 2 Up to 20 characters can be set for the waveform name 3 If the amplitude of the waveform is set for over 5mV it will automatically be scaled to 5mV if not then it will remain the same 4 The waveform automatically offsets to 0mV 5 Support only WhaleTeq format when loaded DFS360 AP supports text file txt only and file content descr...

Page 27: ...ta would be 131 072 number of signals This number must be greater than 1 signal description The description text for the signal sample data The raw data unit is mV Example 250 25000 1 ECG_70 0 324519 0 319712 0 319712 0 319712 0 320913 0 320913 0 320913 0 319712 0 314904 0 313702 0 312500 0 311298 0 307692 0 305288 0 304087 0 301683 0 298077 0 296875 0 296875 0 292067 0 288462 ...

Page 28: ...019 0 259615 0 254808 0 256010 0 254808 0 253606 0 515625 0 519231 0 521635 0 526442 0 531250 0 534856 0 539663 0 543269 0 546875 0 554087 0 560096 0 561298 0 561298 0 561298 0 560096 0 555288 0 551683 0 550481 0 546875 0 542067 0 534856 0 531250 0 524038 0 520433 0 516827 0 514423 ...

Page 29: ... Settings 5 2 PC Software Operation Instructions Note The maximum voltage that the DFS360 DFS360D can receive is 3000V Please confirm the DUT s voltage specification of defibrillation is within 3000V before conducting the test 1 Click Add button at the lower left corner in the main screen to bring up the ECG Output window Select the desired test item from the dropdown list to add to the Test List ...

Page 30: ...m may be cut when using the Child mode and test results may be inaccurate because the shock energy exceeds the upper voltage limit 2 Click the Start button and the DFS360 DFS360D will start sending the ECG signals to the AED defibrillator Users may check the working status of the DFS360 DFS360D by the Test List 3 The AED will start analyzing the ECG signals and determine whether a shock is needed ...

Page 31: ... Item Name Description U Timer If there are more than one test item users need to wait for 30 seconds to start the next one Note The countdown for DFS360D is 15 seconds V Shock Waveform Display the shock waveform gave by the DUT As figure 25 users can move the time cursor and voltage cursor on the x axis U W X V ...

Page 32: ...ergy in Joules Parameters and Characteristic Points Show the energy in Joules parameters and characteristic points of the shock including maximum and minimum values of voltage current time duration and tilt Note Formulas of characteristic points Tilt1 Tilt2 𝑇𝑖𝑙𝑡𝑝ℎ𝑎𝑠𝑒1 𝑉1𝑚𝑎𝑥 𝑉1𝑚𝑖𝑛 𝑉1𝑚𝑎𝑥 Slope1 Slope2 𝑆𝑙𝑜𝑝𝑒𝑝ℎ𝑎𝑠𝑒1 𝑦2 𝑦1 𝑥2 𝑥1 𝑉1𝑚𝑖𝑛 𝑉1𝑚𝑎𝑥 𝑡1𝑚𝑖𝑛 𝑡1𝑚𝑎𝑥 Test results including raw data csv parameters and ...

Page 33: ...e heat within the DFS360 DFS360D is too high the test functions will pause until the temperature drops Figure 26 High Temperature Alert When the temperature in the DFS360 DFS360D is under control the Start button will be unlocked Figure 27 DFS360 DFS360D Return to Normal Temperature High temperature ...

Page 34: ...S360D is turned on the fan will start working for a short while of time power on self test 5 3 4 Over Heat Protection When the temperature in the DFS360 DFS360D is too high the self protection will automatically start working Note The DFS360D has additional hardware over heat protection mechanism which can more effectively prevent device damage 5 4 Optional Software Development Kit SDK WhaleTeq pr...

Page 35: ...st Sequence Conversion to NSR Arrhythmia Convert Combination available in AP Device Identification Product Name Model Number Serial Number Production Year FW version HW version Device Temperature Temperature Value available in SDK Temperature Alert available in AP Overheat Protection available in AP Receive Test Result Biphasic Pulsed Biphasic Waveform V1max Waveform I1max Indirectly Waveform V1av...

Page 36: ...c Lawn Edmark Vmax Imax Indirectly Vreverse Ireverse Indirectly t50 t10 tr 6 Standalone Operation Upon turning on DFS360 DFS360D device and without opening the DFS360 AP on the USB connected computer the DFS360 DFS360D device is in Standalone mode In this mode the ECG pattern that DFS360 DFS360D automatically plays is NSR at 5 mV 80 BPM 7 Calibration and Validation It is recommended to calibrate D...

Page 37: ...send the DFS360 DFS360D back to WhaleTeq for a detailed check For contact information please see section 13 9 Cautions 1 Please use USB 3 0 as power supply for USB Type C port 2 Ensure the USB cable and the AED cable are not entangled with each other Please refer to sec 3 5 1 for detailed information 3 Ensure good airflow around the ventilation holes of the DFS360 DFS360D 4 Follow color code and o...

Page 38: ...ester Main Unit K22 0500101 Open ended AED cable Banana Male 0 5 meter Black K22 0500102 Open ended AED cable Banana Male 0 5 meter Red N61 0210033 Compact Splicing Connector K27 1800304 USB Transfer Cable Type A to Type C Male to Male 1 8 meter Table 11 Ordering info for DFS360D Part No Description Image 300 DFS36D DFS360D AED Defibrillator Tester Dual Main Unit ...

Page 39: ...r K27 1800304 USB Transfer Cable Type A to Type C Male to Male 1 8 meter Optional DFS360 DFS360D Paddle Box Set Table 12 Optional DFS360 DFS360D Paddle Box Set Part No Description Image 300 DFPDOCK DFS Tester Docking Base for Defibrillator E13 070P101 DFS Tester Defibrillation Receiving Plate K22 0100301 AED cable for Paddle Box Female to Male Black K22 0100201 AED cable for Paddle Box Female to M...

Page 40: ... Scan software and barcode scanner Argox barcode scanner model no AS 9400 Optional Calibration Service and Warranty Extension Table 15 Optional Calibration Service and Warranty Extension Part No Description C3 Provides 3 years of calibration service coverage WhaleTeq equipment can be calibrated to original performance on the basis of 1 year interval R3 Extend the limited warranty from 1 year to 3 ...

Page 41: ... 1 year and the number of shocks hasn t reached 15000 30000 times so the one year warranty is adopted 11 Package Contents Table 16 Package Contents for DFS360 Item Q ty DFS360 AED Defibrillator Tester Main Unit 1 Open ended AED cable Banana Male 0 5 meter Black 1 Open ended AED cable Banana Male 0 5 meter Red 1 Compact Splicing Connector 2 USB Cable Type A Male to Type C Male 1 Table 17 Package Co...

Page 42: ...ry Table 18 Revision History Version Description Issue Date 2021 10 22 First Version 2021 10 29 2022 08 08 Update Photo in the cover page 2 3 Energy Measurement Specifications 2 4 ECG Signal Specifications 3 1 Top 3 2 Front 3 3 Rear 3 4 Left and Right Side 3 5 Wiring Diagram 3 5 1 Connection Diagram and Precautions Ch 5 PC Software Operation Ch 10 Ordering Information Ch 11 Package Contents 2022 0...