TROUBLESHOOTING   

PROBLEM(

SOLUTION(S)(

TCU$becomes$very$hot$to$the$touch.$

Discontinue$use$immediately$and$contact$

Westmed$Customer$Service.$

$

Case$components$of$the$TCU$case$become$loose$or$

separate$altogether.$

Retighten$screws$if$all$are$present.$If$some$screws$

are$missing,$contact$Westmed$Customer$Service.$
$

Case$components$of$the$HHT$become$loose$or$

separate$altogether.$

$

Discard$faulty$disposable$unit$and$replace$with$a$

new$HHT.$

Interface$cable$between$the$TCU$and$HHT$becomes$

damaged$from$the$HHT,$or$the$miniBplug$becomes$

damaged.$

$

Discard$faulty$HHT$and$replace$with$a$new$HHT.$

Patient$or$caregiver$unable$to$assemble$parts;$i.e.,$
parts$do$not$seem$to$fit$together$or$mate$properly.$$

$

Refer$to$Instruction$Manual$to$confirm$proper$
parts.$$

Inspect$parts$for$physical$damage$or$deformation$

(e.g.,$outBofBround,$rough$edges).$Replace$parts$

that$have$been$damaged.$

$

Variable$Expiratory$Resistor$for$PEP$and/or$Circulaire$

II$

Hybrid

$for$aerosol$therapy$do$not$seem$to$fit$

properly.$

$

Refer$to$Instruction$Manual$to$confirm$proper$YB

adapter$and$orientation.$$$

Inspect$parts$for$physical$damage$or$deformation$

(e.g.,$outBofBround,$rough$edges).$Replace$parts$

that$have$been$damaged.$
$

Unable$to$attach$nonBWestmed$nebulizer$or$PEP$

device$to$the$Vibralung$Acoustical$Percussor.$

$

Use$only$the$Westmed$Circulaire$II$or$Circulaire$II$

Hybrid

$for$aerosol$therapy$and$Westmed$Variable$

Expiratory$Resistor$for$PEP.$

$
$

!

CARE,   CLEANING   AND   DISPOSAL   

The   TCU   should   undergo   surface   disinfecEon   on   a   daily   basis*   to   minimize   surface   contaminaEon.       Clean   all   external   

surfaces   of   the   TCU   with   a   pre-­‐moistened   germicidal   wipe   product   such   as   Clorox®   DisinfecEng   Wipes.      Be   sure   to   allow   

the   appropriate   contact   Eme   as   specified   by   the   product's   instrucEons.      Ensure   that   moisture   does   not   enter   the   case   of   

the   TCU   at   the   seams   or   at   the   two   electrical   jacks.       The   control   buYons   on   the   faceplate   are   sealed   against   moisture   

entry.   
   
The   HHT   should   undergo   cleaning   and   disinfecEon   on   a   daily   basis*   to   remove   possible   accumulaEon   of   secreEons   and   

to    minimize    surface    contaminaEon.        To    clean    the    HHT,    first    remove    the    Y-­‐adapter,    mouthpiece,    variable    expiratory   

resistor   and   screw-­‐on   cone.       Wash   these   parts   with   soap   and   warm   water,   rinse   with   sterile   water   then   disinfect   by   

soaking   in   a   soluEon   of   1   part   household   bleach   and   50   parts   water   for   3   minutes.       Spread   all   parts   on   a   clean   paper   

towel   to   air   dry.      Also   remove   the   Circulaire,   if   present,   and   clean   it   according   to   its   instrucEons.         
   
Do   not   immerse   the   handle   or   speaker   part   of   the   HHT   in   any   liquid.       Clean   the   outside   surface   of   the   HHT   and   the   

transducer    diaphragm    with    a    pre-­‐moistened    germicidal    wipe    product    such    as    Clorox®    DisinfecEng    Wipes.        Allow    the   

appropriate   contact   Eme   as   specified   by   the   product's   instrucEons.   Allow   to   dry   before   reassembling   and   using.   
   
The   disposable   HHT   and   all   associated   plasEc   parts   may   be   discarded   into   the   usual   waste   stream   in   accord   with      

local   regulaEons.      The   TCU   contains   a      lithium-­‐ion   baYery   pack,   which,   in   most   states,   can   be   disposed   of   via   normal   

waste   management   procedures.      If   in   doubt,   please   check   your   local   regulaEons.   
   

            

   

*   Westmed   recommends   cleaning   and   disinfecEon   at   the   end   of   each   treatment   day.

   

Summary of Contents for Vibralung Acoustical Percussor

Page 1: ...Contraindica ons ACT or use of the Vibralung AcousEcal Percussor especially with PosiEve Expiratory Pressure may be contraindicated in paEents who have untreated air leaks tension pneumothorax bronch...

Page 2: ...gs during each treatment Do not connect or interface the Vibralung AcousEcal Percussor to any airway clearance device or aerosol delivery system other than the Westmed Circulaire II or Circulaire II H...

Page 3: ...andard 120 volt electrical outlet and plug the end of the baYery charger cable into the boYom jack on the len side of the TCU The BaYery Charger plug will not t in the top jack on the len side of the...

Page 4: ...n or compressed air Adjust the opening of the Variable Expiratory Resistor to apply PEP See the Circulaire II or Circulaire II Hybrid package insert for further instrucEons for use Bird s Eye View HHT...

Page 5: ...as well as a Start Pause funcEon and Power On O AddiEonally two LED indicators Charging and Low BaYery illuminate as needed to show the status of the internal baYery Page 5 Connect HHT here Connect Ba...

Page 6: ...de 3 Resume Terminate Repeat To resume an automaEcally terminated session or repeat a session with the same sepngs or start a di erent session with di erent sepngs repeat steps 1 and 2 above and selec...

Page 7: ...ng the Circulaire II Hybrid to the HHT as shown in the boYom gure on page 4 The Circulaire II Hybrid when aYached to the Vibralung AcousEcal Percussor with the Aerosol Y adapter provides a direct path...

Page 8: ...Page 8 TROUBLESHOOTING conEnued next page...

Page 9: ...o minimize surface contaminaEon Clean all external surfaces of the TCU with a pre moistened germicidal wipe product such as Clorox DisinfecEng Wipes Be sure to allow the appropriate contact Eme as spe...

Page 10: ...Page 10 TECHNICAL SPECIFICATIONS conEnued next page...

Page 11: ...Page 11 TECHNICAL SPECIFICATIONS...

Page 12: ...Page 12 ELECTROMAGNETIC EMISSIONS AND ENVIRONMENT...

Page 13: ...placing its internal baYeries or replacing the enEre TCU To iniEate a service request please contact the Westmed Customer Service Department and request a Returned Goods AuthorizaEon RGA Number prior...

Page 14: ...mark of Westmed Inc Circulaire II Hybrid is a trademark of Westmed Inc PN 71944 Rev 01 Westmed Inc 5580 S Nogales Highway Tucson AZ 85706 800 975 7987 www westmedinc com For further informa on scan th...

Reviews: