Westmed Vibralung Acoustical Percussor Instructions For Use Manual Download Page 4

If   simultaneous   aerosol   therapy   is   required,   obtain   and   aYach   the   Circulaire   II   

Hybrid

   aerosol   delivery   system   as   shown   

below.      Remove   the   Standard   Y-­‐adapter   (E),   replace   with   the   Aerosol   Y-­‐adapter   (F),   and   transfer   the   mouthpiece   to   the   

Aerosol    Y-­‐adapter.        Insert    the    Circulaire    II   

Hybrid

    aerosol    delivery    system    on    to    the    tail    of    the    Aerosol    Y-­‐adapter    as   

shown.   The   outlet   port   of   the   cone   should   be   at   the   3   o’clock   posiEon   as   you   are   facing   it,   and   the   assembly   should   be   

aligned   as   shown   in   the   ‘Bird’s   Eye   View’   below:

   

   

Operate   the   nebulizer   with   oxygen   or   compressed   air.      Adjust   the   opening   of   the   Variable   Expiratory   Resistor   to   apply   

PEP.      See   the   Circulaire   II      or   Circulaire   II   

Hybrid

   package   insert   for   further   instrucEons   for   use.      

   

Bird’s   Eye   View   

HHT   ASSEMBLY   

WITH

   AEROSOL   

Page   4   

   HHT   ASSEMBLY   

WITHOUT

   AEROSOL   

The    assembled    Vibralung    AcousEcal    Percussor    Hand-­‐held    Transducer    (HHT)    is    shown    below.    This    assembly    uses    the   

Standard   Y-­‐adapter   (E).      Connect   the   long   end   of   the   Standard   Y-­‐adapter   to   the   cone.      The   mouthpiece   should   be   at   the   

other   end   and   the   Variable   Expiratory   Resistor   should   be   on   the   tail   of   the   Y.       The   outlet   port   of   the   cone   should   be   at   

the   3   o’clock   posiEon   as   you   are   facing   it,   and   the   assembly   should   be   aligned   as   shown   in   the   “user’s   view”   below:

   

User’s   View,   

“3   o’clock”   alignment   

Summary of Contents for Vibralung Acoustical Percussor

Page 1: ...Contraindica ons ACT or use of the Vibralung AcousEcal Percussor especially with PosiEve Expiratory Pressure may be contraindicated in paEents who have untreated air leaks tension pneumothorax bronch...

Page 2: ...gs during each treatment Do not connect or interface the Vibralung AcousEcal Percussor to any airway clearance device or aerosol delivery system other than the Westmed Circulaire II or Circulaire II H...

Page 3: ...andard 120 volt electrical outlet and plug the end of the baYery charger cable into the boYom jack on the len side of the TCU The BaYery Charger plug will not t in the top jack on the len side of the...

Page 4: ...n or compressed air Adjust the opening of the Variable Expiratory Resistor to apply PEP See the Circulaire II or Circulaire II Hybrid package insert for further instrucEons for use Bird s Eye View HHT...

Page 5: ...as well as a Start Pause funcEon and Power On O AddiEonally two LED indicators Charging and Low BaYery illuminate as needed to show the status of the internal baYery Page 5 Connect HHT here Connect Ba...

Page 6: ...de 3 Resume Terminate Repeat To resume an automaEcally terminated session or repeat a session with the same sepngs or start a di erent session with di erent sepngs repeat steps 1 and 2 above and selec...

Page 7: ...ng the Circulaire II Hybrid to the HHT as shown in the boYom gure on page 4 The Circulaire II Hybrid when aYached to the Vibralung AcousEcal Percussor with the Aerosol Y adapter provides a direct path...

Page 8: ...Page 8 TROUBLESHOOTING conEnued next page...

Page 9: ...o minimize surface contaminaEon Clean all external surfaces of the TCU with a pre moistened germicidal wipe product such as Clorox DisinfecEng Wipes Be sure to allow the appropriate contact Eme as spe...

Page 10: ...Page 10 TECHNICAL SPECIFICATIONS conEnued next page...

Page 11: ...Page 11 TECHNICAL SPECIFICATIONS...

Page 12: ...Page 12 ELECTROMAGNETIC EMISSIONS AND ENVIRONMENT...

Page 13: ...placing its internal baYeries or replacing the enEre TCU To iniEate a service request please contact the Westmed Customer Service Department and request a Returned Goods AuthorizaEon RGA Number prior...

Page 14: ...mark of Westmed Inc Circulaire II Hybrid is a trademark of Westmed Inc PN 71944 Rev 01 Westmed Inc 5580 S Nogales Highway Tucson AZ 85706 800 975 7987 www westmedinc com For further informa on scan th...