Westmed Vibralung Acoustical Percussor Instructions For Use Manual Download Page 6

OPERATING   INSTRUCTIONS   

Refer    to    physician’s    prescrip(on    or    caregivers    instruc(ons    for    appropriate    seZngs    for    the    Vibralung    Acous(cal   

Percussor.       Regardless   of   which   Y-­‐adapter   is   used,   opera(on   of   the   Vibralung   Acous(c   Percussor   is   the   same.       The   

pa(ent   should   hold   the   device   in   one   or   both   hands   while   inser(ng   the   mouthpiece   into   the   mouth   and   breathing   

normally   throughout   the   treatment.         

   

1.

   Power.

      To   power   on   the   TCU,   disconnect   it   from   the   charger,   plug   the   HHT   into   it,   and   then   press   

   the   Power   On/Off   buYon.   

                  •      The   LEDs   embedded   in   the   Power   and   Start/Pause   buYons   illuminate.         

                  •      The   Low   BaYery   LED   illuminates   if   the   baYery   is   in   need   of   recharging.   

                  •      The   Unit   is   in   Pause   mode   upon   powering   on.   

2.

Mode   Selec(on.

      To   select   and   acEvate   a   sepng,      

                  •      Press   one   of   the   5   sepng   buYons   in   the   top   row   (L,   M,   H,   R2   or   R5).   

                                       -­‐   The   LED   embedded   in   the   selected   sepng   buYon   illuminates.   

   

                  •      Press   the   Pause   buYon   to   acEvate   the   device   and   emit   sound   at   the   selected   sepng.      

                                       -­‐   The   LED   embedded   in   the   Pause   buYon   exEnguishes.   

   

                                          •      Unless   stopped   or   paused,   the   unit   will   operate   at   the   selected   sepng   

                                                         for   10   minute   and   then   terminate.   

   

                                          •      Press   the   Pause   buYon   at   any   Eme   while   acEvated   to   pause   the   session.         

                                                            -­‐   The   LED   embedded   in   the   Pause   buYon   illuminates.   

   

                                          •      Press   the   Pause   buYon   while   paused   to   resume   the   session.         

                                                               -­‐   The   LED   embedded   in   the   Pause   buYon   exEnguishes.   

   

                                          •      If   a   paused   session   is   not   resumed   within   2   minutes   by   pressing   the   Pause   buYon   again,      

                                                      the   session   automaEcally   terminates.   

   

   Important:   If   a   different   treatment   mode   is   desired   before   the   currently   selected   mode   has   terminated,   it   is   

                                                necessary   to   power   the   TCU   off,   then   power   it   on   again   and   select   the   desired   mode.

   

   

   

3.      Resume,   Terminate,   Repeat.

      To   resume   an   automaEcally   terminated   session,   or   repeat   a   session   with   the   same   

                     sepngs,   or   start   a   different   session   with   different   sepngs,   repeat   steps   1   and   2   above   and   select   the   appropriate   

                     sepngs.   

   

   

4.      Treatment.

      To   take   the   treatment,   the   paEent   should   place   the   mouthpiece   in   his   or   her   mouth   and   seal   their   lips      

                     around   it   just   Eghtly   enough   to   maintain   a   comfortable   seal.      The   teeth   may   rest   gently   on   the   mouthpiece   but   the      

                     paEent   should   be   cauEoned   to   not   bite   down   or   chew   on   the   son   plasEc   mouthpiece.   

   

5.      Pa(ent   Posi(on.      

Avoid   recumbent   paEent   posiEoning   while   using   the   Vibralung   AcousEcal   Percussor.      PaEent   

                  should   be   sipng   upright   or   at   no   more   than   a   30°   recumbent   angle   to   facilitate   coughing   and   expectoraEon.   

   

6.      Breathing   PaQern.

      The   paEent   should   breathe   normally   during   the   treatment.      As   with   any   breathing   treatment,      

                  hypervenElaEon   or   excessively   deep   or   fast   breathing   may   result   in   dizziness   or   light-­‐headedness.      The   treatment   

                  may   be   paused   at   any   Eme   and   resumed   if   the   paEent   needs   to   rest.      

   

7.         Posi(ve   Expiratory   Pressure   (PEP)   Therapy.

      Adjust   the   Variable   Expiratory   Resistor   to   apply   PEP   during   resEng   

                     breathing.      Minimal   PEP   is   applied   when   the   orifice   is   wide   open   (60°)   and   maximal   PEP   when   the   orifice   is   at   the   

                     minimal   posiEon   (10°).      An   opEonal   PEP   manometer   is   available   so   that   PEP   levels   can   be   monitored.      Minimal   PEP   

                     levels   will   be   observed   during   quiet   Edal   breathing   due   to   relaEvely   low   respiratory   flowrates.      If   desired,   the   

                     paEent   may   be   instructed   to   perform   an   extended   PEP   maneuver   by   taking   a   deep   breath   and   then   exhaling   slowly   

                     while   maintaining   an   expiratory   flowrate   sufficient   to   keep   the   PEP   manometer   in   the   10   to   20   cmH2O   pressure   

                     range   for   3   to   5   seconds.      This   maneuver   can   be   repeated   10   to   20   Emes   or   as   onen   as   desired   for   the   paEent.   

Page   6   

Charging   &   Low   BaQery   LEDs   

Summary of Contents for Vibralung Acoustical Percussor

Page 1: ...Contraindica ons ACT or use of the Vibralung AcousEcal Percussor especially with PosiEve Expiratory Pressure may be contraindicated in paEents who have untreated air leaks tension pneumothorax bronch...

Page 2: ...gs during each treatment Do not connect or interface the Vibralung AcousEcal Percussor to any airway clearance device or aerosol delivery system other than the Westmed Circulaire II or Circulaire II H...

Page 3: ...andard 120 volt electrical outlet and plug the end of the baYery charger cable into the boYom jack on the len side of the TCU The BaYery Charger plug will not t in the top jack on the len side of the...

Page 4: ...n or compressed air Adjust the opening of the Variable Expiratory Resistor to apply PEP See the Circulaire II or Circulaire II Hybrid package insert for further instrucEons for use Bird s Eye View HHT...

Page 5: ...as well as a Start Pause funcEon and Power On O AddiEonally two LED indicators Charging and Low BaYery illuminate as needed to show the status of the internal baYery Page 5 Connect HHT here Connect Ba...

Page 6: ...de 3 Resume Terminate Repeat To resume an automaEcally terminated session or repeat a session with the same sepngs or start a di erent session with di erent sepngs repeat steps 1 and 2 above and selec...

Page 7: ...ng the Circulaire II Hybrid to the HHT as shown in the boYom gure on page 4 The Circulaire II Hybrid when aYached to the Vibralung AcousEcal Percussor with the Aerosol Y adapter provides a direct path...

Page 8: ...Page 8 TROUBLESHOOTING conEnued next page...

Page 9: ...o minimize surface contaminaEon Clean all external surfaces of the TCU with a pre moistened germicidal wipe product such as Clorox DisinfecEng Wipes Be sure to allow the appropriate contact Eme as spe...

Page 10: ...Page 10 TECHNICAL SPECIFICATIONS conEnued next page...

Page 11: ...Page 11 TECHNICAL SPECIFICATIONS...

Page 12: ...Page 12 ELECTROMAGNETIC EMISSIONS AND ENVIRONMENT...

Page 13: ...placing its internal baYeries or replacing the enEre TCU To iniEate a service request please contact the Westmed Customer Service Department and request a Returned Goods AuthorizaEon RGA Number prior...