3. Safety Precautions & Maintenance

3.1 Safety Precautions

1.  This product is designed for laboratory use only. Always follow good laboratory 

practice.

2.  Fill the tank prior to connection to electrical supply.
3.  Use caution when topping up or draining the tank. If this product is not used in 

accordance with these instructions, then basic safety protection may be affected.

4.  The mains supply cord fitted to this product is heat resistant and should be 

replaced with an equivalent type by a qualified electrician.

5.  Ensure that the power supply has a safety earth (ground) terminal. 
6.  Ensure that the mains switch, and power supply connector are accessible during use.
7.  Before using any cleaning or decontamination method please refer to the 

Maintenance section to ensure the proposed method will not damage the unit. 

8.  Connect only to a power supply with the corresponding voltage to that specified on 

the rating label positioned on the rear of the unit.

9.  Do not block ventilation slots during use and always follow installation instructions.
10.  Ensure substances being heated present no risk of a hazard (explosion, implosion,  

or release of toxic or flammable gases) or that these have been addressed. When 

heating substances where liberation of gases occurs, suitable extraction should  

be used.

11.  Use only liquids specified within this Instruction Manual within their specified 

temperature range. 

12.  Drain all liquid before moving the water bath. 
13.  We recommend using a lid above 60˚C. Take care when lifting the lid as steam and  

hot vapours can cause scolding.

14.  Use a thermometer to check the temperature - do not touch the liquid.

Liquid Levels:

Always ensure that the water bath is disconnected from the electrical supply before 

filling or emptying the unit.

Tank Capacity

Minimum Liquid Level

Maximum Liquid Level

14 & 22L

70mm above the shelf

Not exceeding the ridge  

in the tank

28L

130mm above the shelf

Not exceeding the ridge  

in the tank