Timer Operation:
1. To start the timer, press the RUN key for more than 1.5 seconds. The timer will
automatically start once the bath has reached set temperature. Heater indicator will
illuminate continuously while liquid temperature is being raised to set point.
2. Heater indicator will flash when set temperature has been reached and is being
maintained. Timer indicator will flash showing the timer is running.
3. To view time remaining, press the UP arrow until the ‘t’ appears and time remaining
is displayed as h:mm.
4. An audible beeping and the ‘END’ message indicates timed period has finished.
Press the RUN key to deactivate the beep and clear the ‘END’ message.
Power Interrupted During Timer Mode:
If the power is interrupted during the timer mode, the display shows ‘P.OFF’ when resumed.
To clear press and hold the RUN key until display reverts to actual temperature. Timer mode
will then continue. To deactivate the timer press and hold the RUN key.
Under & Over Temperature Alarm:
The under/over temperature alarm is automatically set 4°C below/above set temperature.
When in alarm condition the ‘Under/Over Temperature Alarm’ warning indicator illuminates,
and actual bath temperature is shown. All heating/cooling is switched off. Note: if an accessory
dip cooler is being used cooling will continue. Once water temperature has risen above/fallen
below alarm setting then the indicator clears, and actual bath temperature is displayed.
NOTE: The motor remains on, while the heater is switched off.
Low Liquid Level Float Switch ‘FILL’ Message:
The water bath features a low liquid level alarm - display will read ‘FILL’. Heating will be
switched off if the liquid falls below the recommended level. Stirring action will continue.
To reset top up the liquid carefully as spitting may occur.
5. Product Disposal
In the case that the product is to be disposed of, the relevant legal regulations are to
be observed.