15
a)
C1 = sensitivity setting -values: 0 1 2 3 4
The higher the value, the lower the sensitivity and higher stability.
Filter 0: Setting for dispensing.
Filter 1/2: Sensitive and fast, very quiet using location.
Filter 3/4: Robust but slow, busy using location.
Default setting is 2 or 1.
b)
C2 = filtering strength setting -values: 0 1 2 3
The higher the value, the lower the sensitivity and higher stability.
Filter 0: Setting for dispensing.
Filter 1/2: Sensitive and fast, very quiet using location.
Filter 3: Robust but slow, busy using location.
Default setting is 2 or 1.
c)
C3 = serial port setting - transmission speed (RS-232)
2(600), 3(1200), 4(2400), 5(4800), 6(9600) -Default setting is 2.
d)
C4 = device number / communication setting (RS-232)
Press [
TARE
] button changes the value smaller. Press [ ] button changes the value bigger.
The number corresponds to the first data signal to the command control of the computer.
Default setting is 27 (1B
hex
).
e)
C5 = setting the mode for the backlight
0(always on),1(automatic after the weight has changed),2(usually off) - Default setting is 0.
f)
C6 = setting the internal calibration function
0(Internal calibration function on),
1(Internal calibration function off, press the “CAL” key the internal calibration will not work),
2(change the “CAL” key as the external calibration key, press “CAL” key the balance begin the
external calibration mode) –
Default setting is 0.
g)
Save settings
After all points have been configured, press [CAL] button returns the balance automatically into
weighing mode.
h)
Restore the factory settings
In the off state press the [
TARE
] button and press the [ ] button turn on the balance, keep the
[
TARE
] button pressed until the balance returns automatically into weighing mode. All the
parameters will be reset to default settings. Please re-calibration the balance before using.