10
6.2
TARE
When the pan is empty and indication is different than zero press [
TARE
] key.
Weighing with tare: Place the weighing box and press the[
TARE
] button. After standstill control
the zero display appears. The weight of the container is now internally saved. Weigh the
material, the net weight will be indicated. The weight of the weighing container will be displayed
as a minus number after removing the weighing container. The tare weight is saved until it is
deleted. Remove the load from the balance and press the [
TARE
] button. The tare procedure can
be repeated as many times as necessary, for example with initial weighing of several
components for a mix (add-on weighing). The limit is reached when the total weighing range
capacity is full.
6.3
21CAL - [calibration with external weight]
The balance has an external calibration function.
The adjustment procedures see chap. "External calibration”.
6.4
UNIT- [multifunctional key]
The [UNIT] - button is a multi-function button.
6.4.1
Weighing unit selection [weighing mode]
Under the weighing mode, use the UNIT key to choose weighing units: g (gram), ct (carat), ozt
(troy ounce), oz (ounce), GN (gran), dwt (pennyweight) and lb (pound).
1 g
= 5 ct
= 0.032150747 ozt
= 0.035273962 oz
= 15.43235835 GN
= 0.643014931 dwt
= 0.0022046226 lb
6.5
Print function [print mode]
The balance can be connected via the RS232C interface directly to a printer or PC.
Press the “PRINT” key to send the weight to the output device.
6.6
Pieces counting function [counting mode]
Before the balance can count parts, it must know the average part weight (i.e. reference).
Proceed by putting on a certain number of the parts to be counted. The balance determines the
total weight and divides it by the number of parts, the so-called reference quantity (1, 10, 20, 50,
or 100). Counting is then carried out on the basis of the calculated average piece weight.
As a rule: The higher the reference quantity the higher the counting exactness.
Follow these steps:
a)
Place as many pieces to add-up as required by the reference quantity. Press the “COUNT” key,
display “1” that means into the counting mode, Press the [UNIT] button to enter the setting