15
7.7. Output Rate
The output rate indicates how often the measured value is refreshed per second at the output.
Output Rate
Adjust Output Rate
Normal
Fast
3
Back
7
Run
Normal:
The normal output rate is 100/s.
This means that a new measured value is read out every 10 ms.
Fast :
The fast output rate is 330/s.
This means that a new measured value is read out every 3 ms.
Note: If this mode is activated, the OLED display is switched off
as soon as a PROFINET controller goes online. It’s switch back on
automatically when the controller goes offline again.
7.8. Laser
In the menu item “Laser”, the emitted light can be switched on or off.
Laser
Switch transmitted light on or off
On
Off
3
Back
7
Run
ON:
Switch transmitted light on
OFF:
Switch transmitted light off; the Sensor no longer supplies mea-
sured values
7.9. I/O Test
This function manually changes the output of the Sensor. As a result, it possible to test whether the further pro-
cess is working as desired. The test is automatically terminated once you leave the test menu.
I/O Test
Test of the Sensor outputs
Measured value
Statusbits
Warm-up
Signal Strength
Temperature
3
Back
7
Run
Measured value: Default of a measured value in µm
Statusbits:
The number of the status bit to be set can be selected by
pressing the “+” or “–” button. (see list of Statusbits)
Warm-up:
Default of the warm-up on “ok” or “warm-up”
Signal strength: Default of the signal strength on “ok”, “too low” or “too high”
Temperature:
Default of the temperature on “ok”, “too low” or “too high”
If the Sensor returns to the display view after 30 seconds without pressing the button while the test is still active,
this is indicated by a
in the display view.