wellsystem SPA: EMC information
7.1.1
Table 201: Guidelines and manufacturer's
declaration – Electromagnetic emissions
The wellsystem SPA COMPLETE massage device has been designed for use
in the electromagnetic environment specified below. The device operator
must ensure that it is used such an environment.
Emission measurements
Conformity Electromagnetic environment – Guidelines
HF emissions as per CISPR 11 Group 1
wellsystem SPA COMPLETE uses HF power solely for internal func-
tions. Therefore the HF emissions are very low and it is highly unlikely
that it will cause any interference to adjacent electronic equipment.
HF emissions as per CISPR 11 Class B
wellsystem SPA COMPLETE has been designed for use in all devices,
including those in residential areas, and it has been determined that it
can be used for residential purposes as soon as it is connected up to a
public mains supply that also supplies buildings.
Harmonics
IEC 61000-3-2
Class A
Voltage fluctuations /
Flicker
IEC 61000-3-11
Fulfilled
The obligations listed in Sub-section 6.2.2 of EN 61000-3-11:2000 must
be taken into consideration. In particular, the 0.14 ohm line impedance
must be guaranteed for pulse massaging.