This product is for use on a nominal 120V circuit, and has a grounding plug that looks like the 
plug illustrated in Sketch A in Figure #1. A temporary adapter, which looks like the adapter 
illustrated in sketches Band C, may be used to connect this plug to a 2-pole receptacle as 
shown in sketch B if a properly grounded outlet is not available.

 

The temporary adapter should be used only until a properly grounded outlet (Sketch A) can 
be installed by a qualifi ed electrician. The green colored rigid ear, lug, etc. extending from the 
adapter must be connected to a permanent ground such as a properly grounded outlet box 
cover. Whenever the adapter is used, it must be held in place by a metal screw.

2.25 Extension 

Cords

 

If necessary to use an extension cord, use a 3-wire extension cord that has a 3-blade 
grounding plug, and a 3-slot receptacle that will accept the plug on the product. Make certain 
your extension cord is in good condition. Make certain your extension cord wire size is not less 
than 18 gauge for 25 feet, 16 gauge for 50 feet, and 14 gauge for 100 feet. An undersized cord 
will cause a drop in line voltage resulting in loss of power and overheating. If in doubt, use the 
next heavier cord. The smaller the gauge number, the heavier the cord.

2.26  Dial Vacuum Gauge
 

A dial vacuum gauge is incorporated in the vacuum system. The dial gauge is located in the 
top front section of the system. The dial vacuum gauge is commonly used to indicate vacuum 
level when operating in the rough vacuum range (0 to 29.9 InHg) Dial vacuum gauges give 
negative pressure - that is pressure below atmospheric. The reference point for the gauges 
is atmospheric pressure. Please keep in mind that atmospheric pressure tends to vary from 
day to day. As a result of this variability, a dial vacuum gauge will indicate slightly different 
maximum vacuum readings from day to day.

2.27  Vacuum Level Regulation
 

Vacuum level regulation is accomplished with an adjustable bleed valve. The adjustable bleed 
valve allows close control of pressure in your apparatus. Pressure may be set between roughly 
atmospheric pressure and the maximum vacuum obtainable (9 torr, 29.6 InHg).

2.28 Vacuum 

Connections

 

All Dry Vacuum Pumps come with intake and discharge hose barbs which accept 3/8” ID 
rubber pressure/vacuum hose. Hose clamps should be used to hold the hose in place.  Since 
all six models operate in the viscous fl ow range, the small diameter of the hose will generate 
minimal conductance loss. For best results, Welch recommends the length of the tubing 
between the pump and the chamber be kept as small as possible.

2.29 Vacuum 

Gauges

 

Typical in the rough vacuum range, a dial vacuum gauge is used to measure pressure in mm, 
Hg or inches of Hg. The dial vacuum gauge gives negative pressure – that is pressure below 
atmospheric. The reference point for the vacuum gauge is atmospheric pressure. Please keep 
in mind that atmospheric pressure tends to vary from day to day. As a result of variability, the 
dial vacuum gauge will indicate slightly different maximum vacuum readings from day to day.

Summary of Contents for 2562B-01

Page 1: ...r organic vapors or gases Warning Be sure to properly identify intake and discharge before using pump See section 2 5 Part No 642518D Printed in USA Welch Ilmvac 5621 W Howard Street Niles IL 60714 Ph...

Page 2: ...ons are found where failure to observe the instruction could result in damage to the equipment associated equipment and process These units conform to the SI International system of units of measureme...

Page 3: ...Specifications 3 01 Specification Chart Section 04 Motor Power Specifications Features 4 01 Motor Specification Chart 4 02 Changing the Voltage Settings for Global Motor Option on A Models 4 03 Changi...

Page 4: ...e potential of the pump head bursting and causing possible injury to personnel in the area I 17 All electrical products generate heat To avoid serious burns never touch unit during or immediately afte...

Page 5: ...e contact The pump should be located as closely to its system as possible in order to utilize it most efficiently The pump is rated for indoor use only Maximum altitude 2000 meters Operating temperatu...

Page 6: ...Identification Symbols Power Off Power On 2 24 Grounding Instructions This product should be grounded In the event of an electrical short circuit grounding reduces the risk of electric shock by provi...

Page 7: ...nly used to indicate vacuum level when operating in the rough vacuum range 0 to 29 9 InHg Dial vacuum gauges give negative pressure that is pressure below atmospheric The reference point for the gauge...

Page 8: ...is if you have droplets coming out of the exhaust port If you see droplets of water Welch recommends you use a cold trap to capture the water before entering the pump or dilute the water vapor stream...

Page 9: ...3 13 Leak Detection The importance of eliminating all leaks in a vacuum system is obvious The pump must remove this added volume of leaked gas to maintain the desired vacuum Leaks for these pumps can...

Page 10: ...d A simple criterion for the condition of the pump is the determination of its maximum vacuum capability This can be accomplished by blocking of the intake and reading the vacuum level on the gauge Se...

Page 11: ...NPT 3 8 Weight lbs 16 5 Kg 7 5 Overall Dimensions L in Inches cm 17 0 43 2 W in Inches cm 7 5 19 1 H in Inches cm 12 0 30 5 Shipping Weight lbs 24 Kg 10 9 Shipping Dimensions L in Inches 21 5 W in Inc...

Page 12: ...12 6 11 Pumping Speed Curves...

Page 13: ...6 12 Dimensional Drawing 244 0 CORD ASSY 8 25 4X 8 32 UNC 2B THRU 7 09 6 67 9 45 INTAKE NIPPLE FOR 3 8 TUBING MUFFLER 11 67 OUTLET INLET 182 6 3 50 5 38 7 19 296 4 240 0 209 5 9 61 180 0 169 4 88 9 13...

Page 14: ...1 2 666544 Connecting Rod Ass y 2 2 2 621755 502 Valve Plate Ass y 2 3 2 623071 O Ring Cylinder 2 2 4 2 623143 O Ring Valve Plate 2 2 5 2 641255 Exhaust Muffler 2 1 6 6 625175 Head Screw 6 7 2 624496...

Page 15: ...An inlet trap to help prevent ingestion of fluids into the pump a dial gauge for continuous vacuum level monitoring and a vacuum regulator to adjust vacuum levels Just plug it in and you re up and ru...

Page 16: ...pressed or implied The method of executing this warranty servicing adjusting repairing or replacing shall be at the discretion of Welch Ilmvac Vacuum pumps that have been operated within a vacuum syst...

Reviews: