Page | 6
ELECTROMAGNETIC COMPATIBILITY (EMC)
NOTE: This section applies to the WAM PC and the UTK devices when connected to a computer. This
section does not apply to the computer alone.
Electromagnetic compatibility with surrounding devices should be assessed when using the device.
An electronic device can either generate or receive electromagnetic interference. Testing for
electromagnetic compatibility (EMC) has been performed on the device according to the international
standard for EMC for medical devices (IEC 60601-1-2). This IEC standard has been adopted in Europe
as the European Norm (EN 60601-1-2).
The device should not be used adjacent to, or stacked on top of other equipment. If the device must be
used adjacent to or stacked on top of other equipment, verify that the device operates in an acceptable
manner in the configuration in which it will be used.
Fixed, portable, and mobile radio frequency communications equipment can affect the performance of
medical equipment. See the table below for recommended separation distances between the radio
equipment and the device.
The use of accessories and cables other than those specified below, may result in increased emissions or
decreased immunity of the device.
Guidance and Manufacturer’s Declaration: Electromagnetic Emissions
The equipment is intended for use in the electromagnetic environment specified in the table below. The
customer or the user of the equipment should ensure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment: Guidance
RF Emissions
CISPR 11
Group 2
The equipment must emit electromagnetic energy in order to
perform its intended function. Nearby electronic equipment may
be affected.
RF Emissions
CISPR 11
Class A
The equipment is suitable for use in all establishments other
than domestic establishments and those directly connected to
the public low-voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic Emissions
IEC 61000-3-2
Complies
Voltage Fluctuations/
Flicker Emissions
IEC 61000-3-3
Complies