INTRODUCTION
Page | 18
Specifications
Feature
Specification*
Instrument Type
12-lead wireless acquisition module with memory for resting
ECG
Input Channels
8 channels: LA
– RA, LL – RA, V
x
– RA (where
x
is 1
– 6)
ECG Leads Acquired
I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, and V6
WAM PC Transmission Protocol
Bidirectional and frequency hopping; beacon and response
method links a single WAM PC to a single UTK
Frequency Range
2400.96 MHz to 2482.56 MHz
Maximum WAM PC / UTK Distance
10 feet / 3 meters
10 Wire Lead Set
RA, LA, RL, LL, V1, V2, V3, V4, V5, and V6 (R, L, N, F, C1, C2,
C3, C4, C5, and C6) with detachable lead wires
Sampling Rate
40,000 samples/second/channel acquisition; 1,000
samples/second/channel analysis and storage
Voltage Resolution
1.875 microvolts
User Interface
Power and acquisition buttons
Feedback with orange/green LED and audible tone for power
status, operating mode, lead fail, and low battery
Defibrillator Protection
Complies with AAMI standards and IEC 60601-2-25
Memory Capacity
60 seconds
Device Classification
Type CF, battery operated
Weight
6.7 oz. (190 g) with battery
Dimensions
4.45 x 4.25 x 1.1” (11.3 x 10.8 x 2.79 cm)
Battery Life
1 AA alkaline battery typically powers WAM PC for acquisition of
250 resting ECGs
*Specifications subject to change without notice.
WAM / UTK
Radio specifications and certification information for the Wireless Acquisition Module (WAM) and USB