Hygienic preparation
29
EN
•
Make sure no liquid gets into the housing. Never immerse the unit in dis-
infectant or any other liquid. This can damage the unit and consequently
endanger users and patients.
Caution!
•
Make sure no liquid gets into the housing via the
pressure measurement port. This may cause the unit
to malfunction.
The housing, power pack and power supply cable of the
SOMNO
comfort
2e are cleaned by being simply wiped
over with disinfectant.
Breathing hose
•
When disinfecting use the same procedure as for cleaning.
•
The breathing hose must not be heated to temperatures in excess of
70 °C. Sterilisation is not permitted.
Accessories
To disinfect or sterilise the exhalation systems and the mask, please see the section
"Disinfection and Sterilisation" in the relevant operating instructions.
5.6 Disinfecting the SOMNO
aqua
The humidifier can be disinfected. Please observe the operating instructions for the disin-
fectant used. We recommend that you wear suitable gloves when disinfecting the equip-
ment (e.g. household or disposable gloves).
•
The procedure for dismantling and assembling the humidifier is the same
as for cleaning.
For hygiene reasons, we recommend that the plastic parts are replaced after a maximum
of two years use.
Pressure measurement port