Operation
19
EN
4. Operation
4.1 Automatic switch on/off
Activate automatic mode
M = manual
A = automatic
1.
With the SOMNO
comfort
2e switched off, press the softstart button until the current
setting "
M
" starts to flash in the display.
2.
Press the on/off switch to change the setting to "
A
".
If a button is not pressed for 5 seconds, the current setting is saved.
When the automatic system is activated, you can switch SOMNO
comfort
2e on and off
using the on/off switch
or
using the pressure available in the mask.
•
SOMNO
comfort
2e is switched on automatically as soon as the patient
breathes into the mask (pressure > 0.5 hPa).
•
If the mask is removed, the unit will switch off after approx. 15 seconds.
Note:
If the JOYCE Full Face mask is used, the device does not switch on/off automatically,
even if setting "A" is selected. The emergency exhalation valve of the mask means that
the device cannot detect the pressure change required for switching.
Switch off automatic system
M = manual
A = automatic
1.
With the SOMNO
comfort
2e switched off, press the softstart button until the current
setting "
A
" starts to flash in the display.
2.
Press the on/off switch to change the setting to "
M
".
If a button is not pressed for 5 seconds, the current setting is saved.
When the automatic system is switched off, you can only switch SOMNO
comfort
2e on
and off using the on/off switch.
If the unit is not used for 15 minutes, it switches off automatically.