Setup of Equipment
17
EN
1.
Open the filling cap of the humidifier.
2.
Fill the humidifier up to the mark "max
" with
distilled water.
If no distilled water is available, low calcium boiled
water can be used instead (but not as a rule!). Check
that the heater rod does not scale up (see “5.4 Des-
caling the SOMNOaqua” on page 28).
3.
Firmly close the humidifier with the filling cap.
4.
After filling, check that the humidifier is not leaking.
Run your finger over the bottom of the humidifier or
place the humidifier on a piece of kitchen paper. Any
escaping water will then be easy to spot.
Adapting the humidifier
1.
If necessary, detach the breathing hose from the ther-
apy unit by pushing in the catch on the adapter. Pull
the breathing hose and adapter off the unit outlet.
2.
Push the cover sideways out of the humidifier stub on
the therapy unit.
3.
Push the filled humidifier as far as it will go into the
stub on the unit.
4.
Insert the breathing hose adapter into the outlet of
the therapy unit. Check that the locking catch is
pointing upwards and has latched into the groove in
the unit outlet.
max
max
H
2
O
max
e
click!