36
Repairguide for VENTImotion/VENTIlogic
7. Dispose of the foam
from the expansion
space and replace box
.
or:
Replace the complete filter holder
8. If you have not replaced the filter holder, insert
space.
9. Connect box and filter holder. To do so, insert
the intake-side hose in both parts. The smooth
part of the hose must be inserted in the filter
holder in each case.
10. Insert the box and the filter holder back in the
device (steps
in reverse sequence). To
facilitate assembly of the intake-side hose
apply a little 70 % isopropanol to the circum-
ference.
11. Refit the mains supply unit (see "7.4 Replace
mains supply unit" on page 33).
12. Refit the power board (see "7.5 Replace
13. Close the device (see "7.3 Close the device"
14. Test the device (see "4. Test the device" on