32
Repairguide for VENTImotion/VENTIlogic
4. Open the bottom part of the housing
away
to the side.
5. Undo the connecting cables for the humidifier
6. You can now put the bottom part of the
housing
to one side.
7.3 Close the device
Tools and equipment required
• ESD protected work-station
• 3 mm Allen key
1. Hold the bottom part of the housing
up to
the side of the top part of the housing
.
2. Plug the connecting cables for the humidifier
and the alarm
3. Check that all hoses and cables are plugged
on firmly. If necessary,
carefully
plug these fully
onto the connections (hoses) or slots (cables).
4. Place the bottom part of the housing
top part of the housing
Ensure that no cables or hoses are trapped or
bent.
5. Now screw the top part of the housing tight us-
6. Then turn the device back over.