25. Recommendation

Please note that the optimal intensity level varies from person to person.

For optimum result, please do not peel the electrode patches while the device is on.

Important

Store these electrode patches in cool areas with temperature under 30°C in a dry and 
ventilated place. Avoid exposure to sunlight. 

If these electrode patches get dry and lose their stickiness, then put a few drops of water on 
them and rub them evenly and gently to rehydrate them.

Using the foot pads and electrode patches at the same time

If you want to use electrode patches and the foot pads at the same time:

Follow the steps for using the Foot Pads

Then follow the steps for using the Electrode Patches. 

Important note

Selected program will be displayed every 5 mins for 10 seconds during the therapy in the 
PROG/TIME display segment.

Information 

The poly fuse has been used in the devices and that is resettable. We uses on board SMD 
1812L150 LF150 Fuse

Note

Camex wellness Limited will make available on request circuit diagrams, component part 
lists, descriptions, calibration instructions, or other information that will assist service person-
nel to repair those parts of the Circulator (Nerve & Muscle Stimulator) that are designed by 
the Camex Wellness Limited as repairable by the service personnel.

26. Accuracy of Controls and Instruments 

Device is to produce strong non-painful pulsating sensations, often accompanied by 
muscle twitching. Suggests that currents with pulse amplitudes (mA) that generate a 
strong, non-painful pulse sensation are critical for response and therefore pulse amplitude 
should be titrated during treatment to maintain this intensity level last 1 to 85 in some 
scenario above >85 its to be sensation are critical and may harm to user. 

Intensity equation:

pulse_voltage_legs = ( 31V + INT_SETTING_LEGS) * 1.2;

pulse_voltage_pads = ( 21V + INT_SETTING_LEGS) * 1.2;

Summary of Contents for CIRCULATOR

Page 1: ...16 Plot Number Sector 2 Akshar Industrial Park Opp Zydus Cadila Pharma Sarkhej Bavla National Highway Moriaya Ahmedabad 382213 Gujarat India Manufactured For COMFORT CLICK LIMITED Number 7 Doctor Cro...

Page 2: ...ation 9 Care Cleaning and maintenance 10 Purchasing consumables 11 Expected service life 12 Disposal of product WEEE 13 Troubleshooting 14 EMC and Electrical standard requirement 8 9 10 11 11 11 12 13...

Page 3: ...irculation To relieve neuropathy pain osteoarthritic pain acute and chronic body pain varicose vein To provide paralytic nervous activation For strengthening body muscle and relieve body muscle pain F...

Page 4: ...ef by exciting sensory nerves and thereby stimulating either the pain gate mechanism and or the opioid system The type of stimulation delivered by the TENS unit aims to excite stimulate the sensory ne...

Page 5: ...ulation over neck there is a chance that using the electrode patches on neck may cause the closure of airway dif culty in breathing Stimulation over the neck could adversely impact hearing or blood pr...

Page 6: ...cables periodically damage Do not spill liquid or water on the device Do not use this device with Electrode Patches Cable Electrode Patches other than those provided by Camex Wellness Limited The per...

Page 7: ...g with any other types of DC USB Cable may damage the device Usage of any other attachment that is not with your charger may result in electric shock or injury to people Shortening the terminal of the...

Page 8: ...tex free Electrode patches Clean the foot pads with a wet clothe and dry cleaning brush The foot pads are latex free Circulator Nerve and muscle stimulator Do not clean the device while charging 10 Pu...

Page 9: ...se everyone is different Electrodes patches cable broken Replace the cable Electrodes patches cable not connected properly Connect the cable properly You are using this device too frequently Keep suf...

Page 10: ...ode patches have been over used or are dry Replace electrode patches Sensation is too strong Setting is too high Press the button to decrease the setting Sensation is too weak Sensation is too low Pre...

Page 11: ...gress of water with harmful effects IPX8 Protection against continuous immersion in water up to 1 meter or as speci ed by user ON power General Warning Sign Country of manufacture To identify the coun...

Page 12: ...e product within the applicable warranty period If there is not a Distributor you must contact Camex Wellness Limited directly 20 Device foot pads and foot intensity The device should be charged for a...

Page 13: ...he device Step 2 Adjust the intensity through foot intensity level buttons Continue to increase the foot intensity at your comfort level Step 3 As the device is set on Auto Timer for 30 mins post 30 m...

Page 14: ...es Peel off the protective membrane from the electrode patches and carefully place on the clean dry uninjured skin Step 2 Turn on the device through the power button of device Step 3 Adjust the intens...

Page 15: ...10 seconds during the therapy in the PROG TIME display segment Information The poly fuse has been used in the devices and that is resettable We uses on board SMD 1812L150 LF150 Fuse Note Camex wellne...

Page 16: ...on Date 20 06 2022 Document No CWL URM 04A Revision No 00 Revision Date 20 06 2022 29 30 EMS Voltage Waveform Foot Body 12 85v 8 58v Section Voltage Range Foot Body 20 120v 12 84v Section Voltage Rang...

Page 17: ...Document No CWL URM 04A Revision No 00 Revision Date 20 06 2022 Document No CWL URM 04A Revision No 00 Revision Date 20 06 2022 31 32...

Reviews: