10

Specifications

Display:
5XP and 15XP: 

3 ½ digit liquid crystal

display (LCD) with a maximum reading of
1999.

35XP: 

3 ¾ digit liquid crystal display

(LCD) with a maximum reading of 3999.

Polarity: Automatic, positive implied,
negative polarity indication.
Overrange: 

0o

 or 

-0o

 is displayed.

Low battery indication: The 

M

symbol is displayed when the battery
voltage drops below the operating level.
Operating environment: 0 °C to 50 °C
at <70 % R.H.
Power: Single standard 9 V battery,
NEDA 1604, JIS 006P, IEC 6F22.
Dimensions: 155 mm (H) ×72 mm (W)
×32 mm (D)
Weight: Approximately 210 g including
battery.
Overload protection: 1000 V dc or
750 V ac

Accessories: 

One pair test leads TL245,

9 V battery (installed),holster, magnet
strap, Users Manual.
Altitude: 6561.7 Feet (2000 M)

 

Safety: Conforms to EN61010-1, Rev-2;
CAT I 1000V, CAT II 600V, CAT III 300V,
class 2 and pollution deg.2;
CSA 22.2 -1010-1.
EMC: Conforms to EN61326-1. This
product complies with requirements of the
following European Community Directives:
89/ 336/ EEC (Electromagnetic
Compatibility) and 73/ 23/ EEC (Low
Voltage) as amended by 93/ 68/ EEC (CE
Marking). However, electrical noise or
intense electromagnetic fields in the vicinity
of the equipment may disturb the
measurement circuit. Measuring
instruments will also respond to unwanted
signals that may be present within the
measurement circuit. Users should exercise
care and take appropriate precautions to
avoid misleading results when making
measurements in the presence of electronic
interference.

Replacement Parts
TL36 

Test Lead Set

TL245 Test Lead Set
FP200  Fuse – 2 A / 1000 V (15XP and

35XP)

FP300 Fuse – 0.315A / 1000V (5XP)
H2-XR Magne-Grip

®

 Holster, clip, magnet,

and strap

XR-TA  Input Adapter for K-type

thermocouple

TP255 K type thermocouple

See www.metermantesttools.com for
detailed specifications of the Meterman
XP digital multimeters.

5XP Electrical Specifications

(at 23°C ± 5°C, < 75% R.H.)
DC VOLTS
Ranges: 200 mV, 2 V, 20 V, 200 V, 1000 V
Accuracy: ± (1.0% rdg + 1 dgt)

AC VOLTS (45 Hz to 500 Hz)
Ranges: 200 mV, 2V, 20 V, 200 V, 750 V
Accuracy: ± (1.5% rdg + 5 dgts)

DC CURRENT
Ranges: 200 µA, 2 mA, 20 mA, 200 mA,
± (1.5% rdg + 1 dgt)

AC CURRENT (45 Hz to 500 Hz)
Ranges: 200 µA, 2 mA, 20 mA, 200 mA,
± (2.0% rdg + 5 dgts)

RESISTANCE
Ranges: 200 

, 2 k

, 20 k

, 200 k

,

2 M

, 20 M

Accuracy: ± (1.0% rdg + 4 dgts) on 200 to
200 k

 ranges ± (1.5% rdg + 4 dgts) on

2 M

 range ± (3.0% rdg + 5 dgts) on

20 M

 range

CONTINUITY
Audible indication: 75 

±

 25 

DIODE TEST
Test current: 1.0 mA (approximate)
Accuracy: ± (1.5% rdg + 3 dgts)
Open circuit volts: 3.0 dc typical

Summary of Contents for XP Series

Page 1: ...ga ON L4Z 1X9 Tel 905 890 7600 Fax 905 89 6866 European Correspondence Address Meterman Test Tools Europe P O Box 1186 5602 BD Eindhoven The Netherlands Correspondence only no repair or replacement av...

Page 2: ...CAT 600V CAT 1000V CAT 300V COM A MAX 1000V 750V Hz Temp V mA 2A 2A mA 400 1000 A A 35XP RANGE NON CONTACT VOLTAGE HOLD V HZ V auto off C OFF F mA 2A A 1832 400 5 6 4 2 1 35XP 3 Document Dimensions 3...

Page 3: ...ontinuity See Figure 6 5 Testing Diodes See Figure 7 5 Measuring Capacitance 35XP only See Figure 8 5 Measuring Temperature 35XP only See Figure 9 6 Measuring Frequency 35XP only See Figure 10 6 Measu...

Page 4: ...equipment etc where only smaller transient overvoltages may occur and not for primary supply lines overhead lines and cable systems Do not exceed the maximum overload limits per function see specifica...

Page 5: ...To return to autoranging press the RANGE button If RANGE still shows on the display autoranging is not available for the selected function Use autorange for all initial measurements Then when appropr...

Page 6: ...See Figure 3 1 Set the Function Switch to a current function A mA or A 2 Select the desired RANGE 5XP only 3 Connect the test leads Red to mA or A Black to COM 4 Turn off power to the circuit being m...

Page 7: ...rge any capacitors that may influence the reading 5 Connect the test probes across the resistance 6 Listen for the tone that indicates continuity 40 Testing Diodes See Figure 7 1 Set the Function Swit...

Page 8: ...t Voltage See Figure 11 1 Range switch may be set to OFF or any function range 2 Test leads are not used for the NCV test 3 Press the NCV button The display goes blank a tone sounds and the green LED...

Page 9: ...AX button for less than 1 second will put the meter into a mode of displaying the maximum or minimum readings Each time the button is pressed the meter will cycle to the next display mode Press the MI...

Page 10: ...5 A 1000V minimum interrupt rating 30 kA 6 3 x 32 mm Meterman FP300 WARRANTY The XP Series of Digital Multimeters are warranted against any defects of material or workmanship within a period of one 1...

Page 11: ...you Additionally in the United States and Canada In Warranty repair and replacement units can also be sent to a Meterman Test Tools Service Center see below for address Non Warranty Repairs and Repla...

Page 12: ...omagnetic fields in the vicinity of the equipment may disturb the measurement circuit Measuring instruments will also respond to unwanted signals that may be present within the measurement circuit Use...

Page 13: ...mA ranges 2 0 rdg 3 dgts on 2 A range AC CURRENT 45 Hz to 500 Hz Ranges 200 A 2000 A 20 mA 200 mA 2 A Accuracy 1 5 rdg 5 dgts on 200 A to 200 mA ranges 2 5 rdg 5 dgts on 2 A range RESISTANCE Ranges 2...

Page 14: ...30 dgts on 4 nF ranges 5 0 rdg 5 dgts on 40 nF and 400 F ranges 5 0 rdg 15 dgts on 4 mF range TEMPERATURE Ranges 20 C to 1000 C 4 F to 1832 F Accuracy 2 0 rdg 4 C 20 C to 10 C 1 0 rdg 3 C 10 C to 200...

Page 15: ...00 1000 A A 35XP RANGE NON CONTACT VOLTAGE HOLD V HZ V auto off C OFF F mA A A 1832 400 2A MAX FUSED CAT 600V CAT 1000V CAT 300V COM A MAX 1000V 750V Hz Temp V mA 2A A mA 400 1000 A A 35XP RANGE NON C...

Page 16: ...0 A A 35XP RANGE NON CONTACT VOLTAGE HOLD V HZ V auto off C OFF F mA A A 1832 400 2A MAX FUSED CAT 600V CAT 1000V CAT 300V COM A MAX 1000V 750V Hz Temp V mA 2A A mA 400 1000 A A 35XP RANGE NON CONTACT...

Page 17: ...0 1000 A A 35XP RANGE NON CONTACT VOLTAGE HOLD V HZ V auto off C OFF F mA A A 1832 400 2A MAX FUSED CAT 600V CAT 1000V CAT 300V COM A MAX 1000V 750V Hz Temp V mA 2A A mA 400 1000 A A 35XP RANGE NON CO...

Page 18: ...0 1000 A A 35XP RANGE NON CONTACT VOLTAGE HOLD V HZ V auto off C OFF F mA A A 1832 400 2A MAX FUSED CAT 600V CAT 1000V CAT 300V COM A MAX 1000V 750V Hz Temp V mA 2A A mA 400 1000 A A 35XP RANGE NON CO...

Page 19: ...17 10 1 3 2 2 4 K 1 9 3 4 5...

Page 20: ...A V V OFF 200 m 2m 20m 1 5V 9V 200 BATT 5XP 2A MAX FUSED CAT 600V CAT 1000V CAT 300V COM A MAX 1000V 750V Hz Temp V mA 2A A mA 400 1000 A A 35XP RANGE HOLD V HZ V auto off C OFF F mA A A 1832 400 11 1...

Page 21: ...19...

Page 22: ...ga ON L4Z 1X9 Tel 905 890 7600 Fax 905 89 6866 European Correspondence Address Meterman Test Tools Europe P O Box 1186 5602 BD Eindhoven The Netherlands Correspondence only no repair or replacement av...

Reviews: