Figure 2-2. Operating Controls, Rear Panel
OPERATION
Operation as a Voltage Controiled Generator
No preparation for operation is required beyond com-
pletion of the initial installation previously stated in
this section. It is recommended that a one-half hour
warm-up period be allowed for the associated equip-
ment to reach a stabilized operating temperature and
for the Model 132 to attain stated accuracies.
The VCG input connector can be used to externally
control the frequency of the generator. If a positive
voltage is applied to the VCG input terminal, the
frequency will increase from the dial setting, A nega-
tive voltage will cause the frequency to decrease from
the dial setting. The VCG range of the Model 132 is
1000:1.
Operation as a Function Generator
1.
2.
Terminate
OUT connector with
2 watt termination.
Select the desired waveform by setting function
1.
2.
3.
4.
5.
Terminate
OUT connector with 50 ohm
2 watt termination.
Select the desired waveform by setting function
selector to
,
,
or
.
Set FREQ HZ selector to desired multiplier.
Connect external voltage source (dc program-
ming or wideband ac signal) to VCG IN con-
nector.
selector to
,
,
or
,
Set frequency dial and FREQ HZ range multi-
plier for desired output frequency.
Depress MODE - FUNC push button.
Select output signal amplitude by setting OUT-
PUT ATTEN (dB) control to appropriate attenu-
ation position and fine adjusting signal to de-
sired amplitude with OUTPUT VERNIER (dB)
control.
A positive or negative dc offset may be applied
to the waveform by setting the DC OFFSET
(Rear Panel) to the desired level. The peak
signal value plus the offset cannot exceed
V
into 50 ohms.
N O T E
VCG input requires 0 to
volts for
operation over full-scale range, but can
withstand many times maximum input.
Set frequency dial as follows:
a.
For frequency modulation with ac input, set
dial for center frequency.
7