System is now ready for you to apply control solution. You do
not need to set a calibration code for this meter.
[3]
Applying Control Solution:
Shake
the control solution bottle. Discard the
first drop of control solution and wipe
the bottle tip clean using a clean tissue
or paper towel. Dispense a second drop
onto a clean surface, such as an unused
resealable plastic bag.
Bring the tip of the test strip to the control solution sample.
The WaveSense Jazz System will automatically detect if a test
is performed with WaveSense Control Solution.
When control solution is applied, the visual fill window on
the WaveSense Jazz Test Strip turns blue. Your meter will
beep once indicating that you should remove the
tip of the WaveSense Jazz Test Strip from the control
solution sample. On the meter display, the droplet
symbol fills to indicate that the control solution is
filling the WaveSense Jazz Test Strip. The scrolling
Control solution should only be used for 90 days after first
opening the bottle or until the expiration date printed on the
label, whichever comes first.
IMPORTANT: Count forward 90 days from the date you
open a new bottle of WaveSense Control Solution. This is
your discard date. Write the discard date on the WaveSense
Control Solution bottle.
Important WaveSense Control Solution information:
[1]
Use only WaveSense Control Solution with your
WaveSense Jazz System.
[2]
Always replace the cap on the control solution bottle
immediately after use.
[3]
Do not add water or any other liquid to control solution.
[4]
Control solution tests should be performed within the system
operating temperature range of 50 °F to 104 °F (10 °C to 40 °C).
How to perform a control solution test:
[1]
Inserting a WaveSense
Jazz Test Strip:
Insert a new
WaveSense Jazz Test Strip into the
meter. The meter will be activated
when the test strip is fully inserted.
[2]
Meter Ready for Test:
The meter now displays the
droplet symbol and the word “Fill”. The WaveSense Jazz
This is the visual fi ll window.
DO NOT PUT CONTROL
SOLUTION HERE.
Apply control solution here.
The entire visual fi ll
window will turn blue.
Droplet
Symbol: