Summary of Contents for DTS-2077

Page 1: ...Calibration Procedure for the DTS 2077 Time Measurement System WAVECREST Corporation...

Page 2: ...utine expression of that philosophy For this reason any current WAVECREST product may differ in some respect from its published description but will always equal or exceed the original design specific...

Page 3: ...s 2 Reference Voltage Check 3 3 0 100MHz Oscillator Check 4 4 0 Self Calibration 4 5 0 Reference Voltage Operational Check 5 6 0 Calibration Signal Outputs 5 7 0 AC Performance Checks 6 8 0 Arming and...

Page 4: ...cable 1 Virtual Instruments or calibration software Notes 1 Use only electronic test equipment with calibrations traceable to NIST standards 2 Refer to the DTS 2077 User s Guide for general operating...

Page 5: ...ower Supply Checks 2 1 Verify the AC power cord is connected to an appropriate power source 110V 60 Hz or 230V 50 Hz Turn on the main power switch at the rear of the system Then turn on the standby sw...

Page 6: ...DTS as shown in Figure 1B Verify the voltage is 5VDC 0 05V If adjustment is required both 5VDC modules must be alternately adjusted at VOUT The two 5VDC modules provide load sharing and must be balan...

Page 7: ...internal calibration sequence by pushing the INT CAL button on the front panel The display will indicate the time remaining to accomplish the internal calibration counting down on the display When th...

Page 8: ...lower readings on the DTS display It may be necessary to measure and set the DC voltage at the DTS input connectors using a T adapter 6 0 Calibration Signal Outputs 6 1 Using the frequency counter an...

Page 9: ...s its calibration sequence the DTS will display CALIBRATED The previous mode and display then returns 7 2 Go to Menu 11 see Figure 2 6 in the User s Guide and place the cursor on the CLEAR field Then...

Page 10: ...the front panel and press FUNC Verify the displayed measurement is 5000 psec 10 psec with JITTER 6 psec 8 3 Set the function generator to output a square wave with a 5 nsec wide duty cycle and a 1V pk...

Page 11: ...ndby switch on the front panel and the main power switch on the rear of the instrument Remove all cables and unplug the AC power cord from the power source and package the accessories that accompanied...

Page 12: ...ke Road Edina MN 55439 Yes please include me when you distribute future new versions of the DTS 2077 Calibration Procedure NAME __________________________________________________________ COMPANY _____...

Page 13: ...adquarters West Coast Office 7275 Bush Lake Road 1735 Technology Drive Suite 400 Edina MN 55439 San Jose CA 95110 612 831 0030 408 436 9000 FAX 612 831 4474 FAX 408 436 9001 Toll Free 1 800 733 7128 1...