DTS-2077 Calibration Procedure

(17 April,  1998)

2

(1-5)

Using the hex wrench, remove the screws from the bottom.  Store cover and screws in a
safe place.

(1-6)

Turn DTS system over, so that the front panel is right side up.  Then connect the AC
power cord to the rear of the DTS.  Ensure that the main DTS power switch and front
panel standby switch are both off.

CAUTION:

With the cover removed for servicing, there are areas of exposed AC voltage at the
power supply terminal blocks and in the area of the fans.  To avoid exposure to
these hazardous voltages, use extreme caution making adjustments to the power
supplies with the power on.

Note:

The DTS-2077 system is most accurate when the system temperature has reached a
stable condition.  The system should be allowed to warm up with the power "ON"
for at least 1 hour prior to beginning the calibration procedure.  The power should
be left "ON" as often as possible throughout the calibration.

2.0

Power Supply Checks

(2-1)

Verify the AC power cord is connected to an appropriate power source (110V/60 Hz) or
(230V/50 Hz).  Turn on the main power switch at the rear of the system.  Then turn on the
standby switch on the front panel.  Verify the display illuminates, both fans are blowing
air OUT of the system, and the system initiates its self test routine.

Make the following measurements referenced to chassis ground and adjust the power
supply outputs if necessary.

Record the following voltage readings on the test data record (before and after
adjustment).

(2-2)

Using the voltmeter, measure the (+OUT) output of the +15VDC power supply (orange
wire) 
at connector J2, shown in Figure 1A.  Verify the voltage is +15VDC ±0.15V.  If
necessary, 15VDC at VOUT on the 

±

15V module.

(2-3)

Using the voltmeter, measure the (-OUT) output of the ±15VDC power supply (blue
wire) 
at connector J2, shown in Figure 1A.  Verify the voltage is -15VDC 

±

0.15V.  If

necessary, adjust -15VDC at OVP on the 

±

15V module.

(2-4)

Using the voltmeter, measure the (+OUT) output of the +24VDC power supply (yellow
wire) 
at connector J2, shown in Figure 1A.  Verify the voltage is +24VDC +0.24V.  If
necessary, adjust the +24VDC supply.

Summary of Contents for DTS-2077

Page 1: ...Calibration Procedure for the DTS 2077 Time Measurement System WAVECREST Corporation...

Page 2: ...utine expression of that philosophy For this reason any current WAVECREST product may differ in some respect from its published description but will always equal or exceed the original design specific...

Page 3: ...s 2 Reference Voltage Check 3 3 0 100MHz Oscillator Check 4 4 0 Self Calibration 4 5 0 Reference Voltage Operational Check 5 6 0 Calibration Signal Outputs 5 7 0 AC Performance Checks 6 8 0 Arming and...

Page 4: ...cable 1 Virtual Instruments or calibration software Notes 1 Use only electronic test equipment with calibrations traceable to NIST standards 2 Refer to the DTS 2077 User s Guide for general operating...

Page 5: ...ower Supply Checks 2 1 Verify the AC power cord is connected to an appropriate power source 110V 60 Hz or 230V 50 Hz Turn on the main power switch at the rear of the system Then turn on the standby sw...

Page 6: ...DTS as shown in Figure 1B Verify the voltage is 5VDC 0 05V If adjustment is required both 5VDC modules must be alternately adjusted at VOUT The two 5VDC modules provide load sharing and must be balan...

Page 7: ...internal calibration sequence by pushing the INT CAL button on the front panel The display will indicate the time remaining to accomplish the internal calibration counting down on the display When th...

Page 8: ...lower readings on the DTS display It may be necessary to measure and set the DC voltage at the DTS input connectors using a T adapter 6 0 Calibration Signal Outputs 6 1 Using the frequency counter an...

Page 9: ...s its calibration sequence the DTS will display CALIBRATED The previous mode and display then returns 7 2 Go to Menu 11 see Figure 2 6 in the User s Guide and place the cursor on the CLEAR field Then...

Page 10: ...the front panel and press FUNC Verify the displayed measurement is 5000 psec 10 psec with JITTER 6 psec 8 3 Set the function generator to output a square wave with a 5 nsec wide duty cycle and a 1V pk...

Page 11: ...ndby switch on the front panel and the main power switch on the rear of the instrument Remove all cables and unplug the AC power cord from the power source and package the accessories that accompanied...

Page 12: ...ke Road Edina MN 55439 Yes please include me when you distribute future new versions of the DTS 2077 Calibration Procedure NAME __________________________________________________________ COMPANY _____...

Page 13: ...adquarters West Coast Office 7275 Bush Lake Road 1735 Technology Drive Suite 400 Edina MN 55439 San Jose CA 95110 612 831 0030 408 436 9000 FAX 612 831 4474 FAX 408 436 9001 Toll Free 1 800 733 7128 1...

Reviews: