Then connect the micro USB cord of the 

MonitEM-Lab

 to the computer. The 

driver should be installed automatically.
If any problems occur with installation of the driver, consult the Appendix 1: 
Problem with drivers Installation.

5.2 Using the software

5.2.1 First connection

We highly recommend connecting the device for the first time using only 

the USB so that you can check the principal parameters, such as the alarm 

configuration.

The program for managing the 

MonitEM-Lab

  is very easy to use. Start the 

program and connect the device’s USB cord (sequence is not important). The 

program itself will detect the device and display the configuration screen.

5.2.2 

Parameters

We can distinguish between different 
parts of the program:

•  General:

 this indicates the time to 

which the device is set along with 
its serial number and the serial 
number of the connected probe.

•  Status:

 gives information on:

•  Type of power supply.
•  Temperature of the system.
•  Firmware version.
•  Relay state.

•  Measurement:

 allows definition of:

•  The interval of recording in 

minutes or seconds.

•  The window averaging: European recommendation is 6 minutes = 360 

seconds. 0 means no average is done.

 

User’s Manual

Electromagnetic Radiation Monitoring Device MonitEM-Lab   

P. 

10

Summary of Contents for MonitEM-Lab

Page 1: ...34 93 320 80 55 www wavecontrol com info wavecontrol com Safety Quality Service V 2 2 USER S MANUAL Electromagnetic Radiation monitoring device MonitEM Lab...

Page 2: ......

Page 3: ...n is correct to the best of Wavecontrol s knowledge it is provided with no guarantees of completeness accuracy usefulness or timeliness and without any warranties of any kind express or implied Waveco...

Page 4: ...s 10 5 2 3 Visual and audible alarms 11 5 2 4 Actions 12 5 2 5 Online mode 12 5 2 6 Offline mode 13 5 2 7 Malfunctions 13 5 3 MonitEM Lab Ethernet configuration 13 5 3 1 Changing the IP of the MonitEM...

Page 5: ...ctured with high quality materials and components which can be recycled and reused When this crossed out wheeled bin symbol is attached to a product it means the product is covered by the European Dir...

Page 6: ...collecting and managing data from the different devices making up the monitoring system The device can also be used without a Control Centre using the MonitEM Lab Reader PC software to control and dow...

Page 7: ...3 1 Components Red light Probe Ethernet Micro SD card Sound ON OFF button Tripod screw ON OFF switch Main supply IEC C14 connector Alarm output Micro USB User s Manual Electromagnetic Radiation Monit...

Page 8: ...ected to a computer Blinks intermittently when in communication mode Blinks rapidly for 5 seconds after initial calls if communication is successful The LED of the SD indicates that it is writing Do n...

Page 9: ...transforms 110 240V AC 50 60Hz into 12V DC 4 2 USB mode This power supply mode can only be used to change the parameters of the device The USB powers the microprocessor and the SD card so that it can...

Page 10: ...MonitEM Lab must terminate a communication with the server or must save some data in the SD card When connected with the USB MonitEM Lab will not shut down even with the switch in OFF until you remov...

Page 11: ...If Windows indicates that it cannot verify the digital signature for the driver continue with the installation anyway User s Manual Electromagnetic Radiation Monitoring Device MonitEM Lab P 9...

Page 12: ...m and connect the device s USB cord sequence is not important The program itself will detect the device and display the configuration screen 5 2 2 Parameters We can distinguish between different parts...

Page 13: ...onnected 5 2 3 Visual and audible alarms You can define 3 parameters Status light that indicates unit is working by flashing from time to time You can define the blink interval in seconds and you can...

Page 14: ...avecontrol Control Centre web server A new screen will open to start the check If the check is successful it will show that the server has accepted the communication 5 2 5 Online mode not compatible w...

Page 15: ...licking the Tools Save Log menu Send these files to your technical support contact Note if the red light status SD card stays on all the time when the USB is connected please disconnect the USB and re...

Page 16: ...ned with the IP found by the MonitEM Reader Note if the browser asks for a user and password leave them blank On this page go to the Network menu and enable Use the following IP configuration with the...

Page 17: ..._SAMPLE W r The unit with external power and probe inserted will take a measurement and send the result When unit is in normal mode MODE N r it will send the last automatic measurement in setup mode M...

Page 18: ...on OK r n Explanation of response parameters VX Axe X voltage VY Axe Y voltage VZ Axe Z voltage FX Axe X field value FY Axe Y field value FZ Axe Z field value F Total field T Temperature B Battery lev...

Page 19: ...ng configuration Port 10001 Slave ID 17 Function code 3 Registers to read the EMF field in V m Address 50512 Hex 0xC550 Length 2 the EMF field value is sent with 2 registers of 16 bits 32 bits Note th...

Page 20: ...ol recommends to calibrate the unit and probe every 2 year 7 3 Update firmware If you have received a new firmware for the MonitEM Lab with the PC software go to Tools Firmware update menu A new windo...

Page 21: ...expiration of the warranty period by sending a Failure Report including Defect Report Equipment data model serial number purchase date Contact data company name and complete contact details This warr...

Page 22: ...r by clicking right 2 Do not allow Windows to find the driver automatically 3 Select Install from a list or specific location 4 Select D Drivers as the location where D is the letter for the USB drive...

Page 23: ...w you can see an example Wavecontrol MonitEM Lab Data Date 10 14 2020 13 11 20 MonitEM Serial 20MT0825 Probe serial 18WP040767 Frequencies 100kHz 8GHz Units V m Date Time Max Average value Max Instant...

Page 24: ...E EN 55011 2011 A1 2011 Industrial scientific and medical equipment Radio frequency disturbance characteristics Limits and methods of measurement UNE EN 61000 3 3 2009 Limitation of voltage changes vo...

Page 25: ...t techniques Immunity to conducted disturbances induced by radio frequency fields UNE EN 61000 4 11 2005 Testing and measurement techniques Voltage dips short interruptions and voltage variations immu...

Page 26: ......

Page 27: ...www wavecontrol com...

Page 28: ...Safety Quality Service 34 93 320 80 55 www wavecontrol com info wavecontrol com...

Reviews: