Warnings
Operating Environment: Charging – 0° to ~45°C; Discharging -20°to ~ 60°C.
Please ensure the unit is stored in compliance with the operating environment requirements and in a safe and
secure location.
When sending the TnP 3PLM for calibration, please turn off the battery isolation switch which is located in
the storage compartment of the tester. If you have any questions about the battery of your TnP 3PLM please
contact Wavecom for further advice.
Thermal Circuit Breaker
The TnP 3PLM uses a 16A rated thermal circuit breaker which will trip after 30 minutes of continual 20A testing.
To avoid tripping the circuit breaker avoid running >16A tests for extended periods of time. If the circuit breaker
trips, wait a short time for the breaker to cool down, then simply flip the breaker back into the on position.