background image

IS-HF-M100+-29293  Rev 4

 

9

 

Calibration Procedures (continued)

1.  Select the calibration function by pressing the 

Calibrate

 icon.

2.  Select the appropriate calibration standard (Formazin or ProCal) and 

then press 

 to save the selection.

3.  The calibration menu will be displayed as shown above. On the 

left-hand side are selections for the range for calibration. Select the 

highest value required for your application.

4.  Ensure standards are indexed as described in Indexing Calibration 

Cuvettes section prior to calibration.

5.  Press the command button. The command button dialogue will 

guide you through the calibration.

Calibrating using ProCal Standards

1.  Press the High Range A button on the left side of the screen.
2.  Insert the High Range A standard.
3.  Press the Command button in the middle of the screen to start 

calibration. A countdown from 30 will appear on the screen. When 

the countdown is complete, the screen will request the High Range 

B.

4.  Press the High Range B button on the left side of the screen.
5.  Insert the High Range B standard. 
6.  Press the Command button to continue the calibration. A 

countdown from 30 will appear on the screen. When the 

countdown is complete, the screen will request the 1000 NTU 

standard.

7.  Press the 1000 NTU button on the left side of the screen.
8.  Insert the 1000 NTU standard.
9.  Press the Command button to continue the calibration. A 

countdown from 60 will appear on the screen. When the 

countdown is complete, the screen will request the 10 NTU 

standard.

10. Press the 10 NTU button on the left side of the screen.

Select 

Range

Command 

button: 

Follow 

directions 

that appear 

here for 

calibration.

Formazin Calibration Screen

ProCal Calibration Screen

11. Insert the 10 NTU standard.
12. Press the Command button to continue the calibration. A 

countdown from 60 will appear on the screen. When the 

countdown is complete, the screen will request the 0.02 NTU 

standard.

13. Press the 0.02 NTU button on the left side of the screen. 
14. Inert the 0.02 NTU standard.
15. Press the Command button to continue the calibration. A 

countdown from 30 will appear on the screen. 

When the countdown is complete, all buttons on the left side of 

the screen should be green and the Command button should say 

Calibration Good.

Calibrating Using Formazin Standards

1.  Press the 4000 button on the left side of the screen.
2.  Insert the 4000 NTU standard.
3.  Press the Command button in the middle of the screen to start 

calibration. A countdown from 30 will appear on the screen. When 

the countdown is complete, the screen will request the 1000 NTU 

standard.

4.  Press the 1000 NTU button on the left side of the screen. 
5.  Insert the 1000 NTU standard. 
6.  Press the Command button to continue the calibration. A 

countdown from 60 will appear on the screen. When the 

countdown is complete, the screen will request the 10 NTU 

standard.

7.  Press the 10 NTU button on the left side of the screen.
8.  Insert the 10 NTU standard.
9.  Press the Command button to continue the calibration. A 

countdown from 60 will appear on the screen. When the 

countdown is complete, the screen will request the 0.02 NTU 

standard.

10. Press the 0.02 NTU button on the left side of the screen.
11. Insert the 0.02 NTU standard.
12. Press the Command button to continue the calibration. A 

countdown from 30 will appear on the screen. 

When the countdown is complete, all buttons on the left side of 

the screen should be green and the Command button should say 

Calibration Good.

Alternate Calibration Ranges

Alternately you could choose to calibrate to 1000 NTU, which requires 

1000 NTU, 10 NTU and 0.02 NTU. Press the 1000 button on the left 

side and start at step 5 of the previous section.
A calibration for only 10 NTU requires only the 10 NTU and 0.02 NTU 

standards. Press the 10 NTU button on the left side and start at step 8 

of the previous section.
For the two alternate ranges, these will still read the full range with a 

reduced accuracy above the range of calibration.

Failed Calibration

In the case of a failed calibration, the last good calibration is restored 

once the user exits to the Home screen. A failure will be evident when 

the calibration screen shows all red.
The instrument can be used with this calibration with a potentially 

reduced accuracy. The bad calibration will be noted in the calibration 

log and as a reminder, a yellow diagnostic banner will show on the 

home screen. The yellow banner indicates that the last calibration had 

failed and that the instrument is using the previous calibration. The only 

way to remove this indication is to successfully complete a calibration.
Check the standards for cleanliness. If everything looks good, attempt 

the calibration again. If the calibration fails after the second attempt, 

the most likely cause is that the standards are bad or the lamp needs 

replacement.

Summary of Contents for M100+ 28060

Page 1: ...en Calibration 6 Set Date Time 6 Delete Samples 6 Data Resolution 6 Offset 7 Device ID 7 Calibration Reminder 7 Calibration Download 7 Firmware Update 7 Reset Defaults 7 About Device 7 Calibrating the M100 8 Calibration Standards 8 Care of ProCal Standards 8 Indexing Calibration Cuvettes 8 Calibration Procedures 8 Calibration Procedures continued 9 Calibrating using ProCal Standards 9 Calibrating ...

Page 2: ...hould not permit anyone to touch this equipment unless they are over 18 years of age are adequately trained and supervised and have read and understood this manual Owners should also ensure that no unauthorized personnel comes in contact with this equipment If this equipment or any of its parts becomes damaged or needs repair stop using the equipment and contact an experienced service individual i...

Page 3: ...lt in death or serious injury This symbol identifies hazards which if not avoided could result in death or serious injury This symbol identifies hazards which if not avoided could result in minor or moderate injury This symbol identifies practices actions or failure to act which could result in property damage or damage to the equipment This pictorial alerts you to the need to read the manual This...

Page 4: ...ons for the M100 are valid only for measurement of static non flowing samples Touch Screen Standard Inserted in Optics USB Connector Unpacking and Inspection of the Instrument Remove all items from packing carton and carefully inspect to ensure that no visible damage has occurred during shipment and that all items listed below have been received If the items received do not match the order please ...

Page 5: ...igned to provide the ability to customize the instrument according to your needs at any time during normal operation The Setup screen has been split into sub menus to facilitate instrument configuration This section describes how to use each of the sub menus to configure the instrument Turning on the Instrument 1 Confirm that the M100 is safely and properly plugged into a suitable power source 2 T...

Page 6: ...ration The screen has been factory adjusted but can be adjusted if the alignment is off The screen may need to be calibrated if new software has been downloaded It is a 5 position setup 1 Press the Setup icon Select Touch Screen Calibration 2 If this calibration fails restart the instrument by unplugging and reinserting the power plug 3 Press the top screen time banner for 5 seconds to reset to fa...

Page 7: ...nsert the USB memory stick that contains the software into the USB port 3 If the firmware revision on the memory stick is older than the firmware currently installed the instrument will ask if you wish to continue 4 Press the to continue with the firmware installation The upload will take several minutes During the initial upload the screen will show busy and screen anomalies may appear until the ...

Page 8: ...ided as general guidelines for proper care of ProCal standards Standards should only be handled by the top black cap or by the very bottom of the cuvette Keep cuvette glass clean of fingerprints and debris Cuvettes can be cleaned with any domestic glass cleaner and the provided microfiber cloth Prevent ProCal standards from freezing This will most likely destroy them Standards under 100 NTU must b...

Page 9: ...on on the left side of the screen 2 Insert the 4000 NTU standard 3 Press the Command button in the middle of the screen to start calibration A countdown from 30 will appear on the screen When the countdown is complete the screen will request the 1000 NTU standard 4 Press the 1000 NTU button on the left side of the screen 5 Insert the 1000 NTU standard 6 Press the Command button to continue the cal...

Page 10: ...ostic Banner On Off Reading List Setup Calibration Trend Home Data Retrieval Data logging can only save 100 points per Sample ID All data is saved in CSV format and is date and time stamped Sample ID Data can be downloaded to a USB memory stick or printed in the Trend screen after pressing Download Print Calibration data can be downloaded to a USB memory stick in the Calibration Download menu by p...

Page 11: ...Toll Free 888 203 7248 Fax 239 454 0694 Email HF Info wattswater com www hfscientific com Routine Maintenance Routine maintenance should include calibration of the instrument See Calibration section in this manual for more information Cleaning the Cuvette Measurement cuvettes used for the sample should be clean and free of marks or scratches Clean the interior and exterior of the cuvettes with a d...

Page 12: ...e cost of repairing or replacing other property which is damaged if this product does not work properly other costs resulting from labor charges delays vandalism negligence fouling caused by foreign material damage from adverse water conditions chemical or any other circumstances over which the Company has no control In addition the Company shall not be responsible for any costs incidental to the ...