Calibration
Do not use the Manual CinchClamp Tool if it has been
damaged, as this may result in an improper cinch.
After repeated use of the Manual CinchClamp Tool, it
may require calibration. A calibration gauge and bar are
provided in the tool pouch. Follow these steps when cali-
brating a Watts Manual CinchClamp Tool.
1. Insert the Calibration Bar into the tool as shown and ratchet
the Manual CinchClamp™ Tool to the closed position and
hold. Use the calibration tool to gauge the jaw gap. Ideally,
the jaw gap should be between 1.5mm and 1.7mm. If the
1.7mm side of the gauge slides into the gap, then the jaw
space is too large. If the 1.5mm side of the gauge cannot
slide into the gap, then the jaw space is too narrow.
2. To adjust the jaw gap, turn the Manual CinchClamp™ Tool
on its side and locate the Phillips screw positioned on the
ratchet arm. Use a standard Phillips screwdriver to loosen
the screw and to adjust the ratchet arm position. Turning the
arm left or right will change the jaw gap. Tighten screw when
desired jaw gap is reached.
Adjust Calibration
Check Calibration
Adjustment screw
Calibration Bar
Gauge
ES-WP-WPCCT-2 1323
© 2013 Watts
USA:
Tel: (978) 688-1811 • Fax: (978) 794-1848 • www.watts.com
Canada:
Tel: (905) 332-4090 • Fax: (905) 332-7068 • www.watts.ca
A Watts Water Technologies Company