If you purchased your pump prior to 9 February 2017
Check your software version before using this the speed limit setting
Check the software version of the “main processor code” by following
the instructions in section 18.6 Help.
If the software version is less than MKS-2.0 then do not use this
setting as an intermittent fault condition can result in the pump
resetting the speed limit to 125 rpm when the power is turned off to
the pump.
If you require this functionality on a software version lower than
MKS-2.0 please use the 4-20 mA calibration method described in
Section 15 or contact Watson-Marlow aftersales department to discuss
other control methods.
If the version is MKS-2.0 or higher, the speed limit setting may be
used.
Reset run hours
Select
reset run hours
from the control settings menu.
Select
RESET
to zero the run hours counter. The run hours counter can be viewed by
pressing
INFO
from your home screen.
m-qdos-allmodels-en-07