3-4
Changing the Mode of Operation
6. Using PEEK tubing equal to 0.004-inch ID, connect port S of the selector
valve to the ESI probe.
Recommendation:
To reduce peak broadening, use 0.004-inch ID tubing
for sample flow rates
1.2 mL/min; use 0.005-inch ID tubing for sample
flow rates
1.2 mL/min.
Requirements:
•
If you are replacing the tubing between the selector valve and the
probe, minimize the length to reduce peak broadening.
•
When cutting the tubing to length, cut it squarely (that is,
perpendicular to its horizontal axis).
Removing the ESI probe
Required material
Chemical-resistant, powder-free gloves
To remove the ESI probe:
1. Prepare the instrument for working on the source (see
2. Disconnect the tubing from the ESI probe.
3. Disconnect the ESI probe cable from the high-voltage connector.
4. Unscrew the probe locking ring.
Warning:
To avoid electric shock, do not use stainless steel tubing
or stainless steel finger-tight screws to connect the selector valve
to the ESI probe; use the PEEK tubing and natural (beige) colored
PEEK finger-tight screws supplied with the instrument.
Warning:
The LC system connections, ESI probe, and source can
be contaminated with biohazardous and/or toxic materials. Always
wear chemical-resistant, powder-free gloves while performing this
procedure.
Warning:
To avoid electric shock, ensure that the instrument is
prepared for working on the source before commencing this procedure.
Summary of Contents for Xevo TQD
Page 12: ...xii...
Page 20: ...xx Table of Contents...
Page 23: ...Uses and compatibility 1 3 Xevo TQD TP03406...
Page 24: ...1 4 Specifications and Operating Modes Xevo TQD with visor up TP03407...
Page 42: ...1 22 Specifications and Operating Modes...
Page 52: ...2 10 Preparing for Operation...
Page 76: ...3 24 Changing the Mode of Operation...
Page 194: ...4 118 Maintenance Procedures...
Page 208: ...A 14 Safety Advisories...
Page 234: ...B 26 External Connections...
Page 238: ...C 4 Materials of Construction and Compliant Solvents...
Page 250: ...Index 8...