Replacing the ion block source heater
4-87
To fit the new IonSABRE II probe heater:
1. Use the probe adjuster knob to adjust the capillary so that it protrudes
approximately 0.5 mm from the end of the probe.
2. Carefully slide the probe heater over the capillary sleeve on the probe
assembly.
3. Fit the probe heater to the probe assembly, ensuring that the heater is
fully seated on the probe assembly.
4. Fit the probe to the instrument (see
).
5. In the Instrument Console, click API
to start the desolvation gas.
Replacing the ion block source heater
Replace the ion block source heater if it fails to heat when the instrument is
pumped down (evacuated).
Required materials
•
Chemical-resistant, powder-free gloves
Caution:
Take great care not to damage the probe heater’s
electrical connections, capillary sleeve, or capillary when fitting
the heater over the capillary sleeve.
Caution:
To avoid damaging the probe heater’s electrical
connections, do not twist the heater when fitting it to the probe
assembly.
Probe heater connections
Capillary sleeve
Capillary
0.5 mm
Summary of Contents for Xevo TQD
Page 12: ...xii...
Page 20: ...xx Table of Contents...
Page 23: ...Uses and compatibility 1 3 Xevo TQD TP03406...
Page 24: ...1 4 Specifications and Operating Modes Xevo TQD with visor up TP03407...
Page 42: ...1 22 Specifications and Operating Modes...
Page 52: ...2 10 Preparing for Operation...
Page 76: ...3 24 Changing the Mode of Operation...
Page 194: ...4 118 Maintenance Procedures...
Page 208: ...A 14 Safety Advisories...
Page 234: ...B 26 External Connections...
Page 238: ...C 4 Materials of Construction and Compliant Solvents...
Page 250: ...Index 8...